SODIMAT - Model 9073
5-3
Command
Option
Function
51
0*
Automatic temperature compensation with a Pt100
1
Manual temperature compensation
61
------
Manual temperature compensation value
101
------
Flow rate calibration in l/h
The flow rate value displayed can be adjusted to correspond to a
reference value. Press ENTER when stable. Adjust digit by digit the
value of the flow rate desired.
201
0
By addition calibration (CARCAL, AUTOCAL)
1
Manual calibration
2*
"Process" calibration
211
0
Utilizing calibration cartridges (CARCAL)
1
Utilizing calibration solutions (AUTOCAL)
215
0*
Deactivates the automatic calibration mode
1
Activates the automatic calibration mode
217
[168*]
Interval between two automatic calibrations
To visualize the
duration before the calibration, enter value "0000". When the
duration is inferior to 100 hours, the instrument displays the delay
before the calibration under the following format : "Hour:Minutes"
(27:04). Otherwise, only the hours are displayed.
220
[7.0*]
Sample pH value to correct the CAR-CAL value (see §6.3.1)
221
0
1 calibration point
1
2 calibration points
223
0*
Low frequency of the calibration pump for 1 point solution calibration.
1
High frequency of the calibration pump for 1 point solution
calibration.
225
[2 ppm*]
Value of the calibration solution concentration
. It is necessary to
validate this command to execute the automatic calculation to
determine the concentration values in C231 and C241. (See
paragragh 6.2.2 to determine the concentration value needed with
your application).
231
-----
Concentration value of addition 1 calculated for 4 l/h and 25
E
C.
241
-----
Concentration value of addition 2 calculated for 4 l/h and 25
E
C.
251
[0.0*]
Drift of the normality voltage to 25 C (in mV).
B
261
[100*]
Slope value at 25 C in % of the theoritical slope (59,16 mV per
B
decade).
271
[20*]
Maximum time for each calibration point in minutes.
After this
time, the measurement is automatically memorized. The user can
interrupt the calibration cycle at any time by pressing
SELECT
or
validate the measurement by pressing ENTER.
411
-----
Temperature calibration
. The temperature value can be tuned to
correspond to a reference value ; the transmitter allows tuning on a
10 C range.
B
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......