11
Instrument is resistant to exposure of radio-
frequency electromagnetic fields:
-
in the frequency range from 80 to
1000 MHz;
-
in the frequency range from 800 to
960 MHz and from 1.4 to 2.5 GHz (in
digital radiophone interference
environment)
Meets the IEC 61000-6-2:2005
Meets the IEC 61000-6-3:2011
Meets the IEC 61326-1:2012
test level 4 (30V)
Instrument is resistant to electrostatic
discharges
air discharge of 8 kV
contact discharge of 6 kV
test level 3
Thresholds of detection of
-
-
(5±1.5)
⋅
10
-5
mg/l,
detection time is less than 10 s
(3±0.9)
⋅
10
-4
mg/l,
detection time is less than 15 s
Indication of the detected CWA
:
-
-
arsenic-containing compound
yellow LEDs bar
red LEDs bar
Instrument case protection class
IP55
Communication with PC
IR link, IrDA standard
USB channel (for PM2012MB)
Weight
no more 0.77 kg
Dimensions
no more 66х47х195 mm