44
CALIPERS
On-Line Calipers are giving inaccurate readings
1.
Re-calibrate as instructed on the screen. (Defaults/Calibration).
2.
Go into a 3-site Skinfold test.
3.
Zero the calipers.
4.
Place the 25.4mm calibration dowel in the jaws of the calipers lengthwise.
5.
Press the “Measure” switch on the calipers.
6.
The reading on the screen should be within 24.5mm and 26.0mm. Do this 3
times and note each reading.
7.
If the readings are not within 24.5mm and 26.0mm, re-calibrate and repeat.
8.
If the dial
on the calipers
do not read exactly 25.4mm when the dowel is in
the calipers, do not worry! The dial on the caliper is not as accurate as the
digital reading being sent to the computer.
9.
If still not within the above values, call the Polar/HealthFirst Corporation
Technical Support Line (505-344-9679).
SIT & REACH (Range of Motion)
Range of Motion handle is not retracting
1.
If the ROM handle is not retracting, the internal mechanism needs to be
repaired by Polar/HealthFirst service personnel.
2.
Remove the ROM unit from the Sit & Reach station using a 5/64” Allen
wrench provided with your system.
3.
Remove the four screws mounting the unit to the station.
4.
Unplug the ROM cable from the TriFIT 620 unit.
5.
Remove the ROM unit.
6.
If out of warranty, you must complete a Service Form before equipment can
be repaired. Call 505-344-1400 ext 107 to request a form or go to
http://www.healthfirstusa.com/support
to download.
Range of Motion is inconsistent
1.
Re-calibrate ROM (Defaults/Calibration) according to the instructions on the
screen.
*It is much easier and more accurate to calibrate the ROM with two people.
2.
Enter into a client test and perform another Sit and Reach test.
3.
When the screen prompts to “Zero the System”, pull the handle out exactly
12” from the foot plate and press the Ent er key.
4.
Return the handle slowly to the foot plate and allow the TriFIT to take the
reading.
5.
The measurement should be within ½ “ of 12”.
6.
If the measurement is not within ½”, re-calibrate and repeat.