32
Range of Motion
1.
Select
Sit & Reach
Click
Calibrate.
The portable 620 Sit & Reach station
calibrates exactly the same as the 600
system.
2.
Pull the handle slowly out to the foot
plate and click the
Next>
button.
3.
Pull the handle out exactly 12” PAST
the footplate*, then click
Next>.
The correct calibration constant is
between 0.45 and 0.60.
*
You must have a 12” ruler when performing
this calibration.
Note
: It is easier and typically more accurate to
use two people when calibrating the Range of
Motion.
WARNING
: Use care when pulling out and returning the handle. Do not pull the handle out
too far and when a client has finished the test be careful they do not let go of the handle and
let it snap back. This is a delicate measurement device.
12”