8. TEST AND ADJUSTMENT FUNCTIONS
8-1. Test and Adjustment Functions
The specifications for inspection and the calibration program are described below.
1) Seven-segment display specifications.
2) To shift to the test/adjustment mode, press the “+” and “-” buttons simultaneously in the sleep mode and
then press the Power button.
The numeric LED display is <AP>.
3) Shift to each test mode by the button operation below.
1 Printer test mode <4. SPC> (*1)
A test print is made from the printer connected to a copy board when you press the Feed/Stop button
and then the print button during button operation.
2 Calibration mode start <2. SPC> (*1)
Calibration is executed when you press the Print button during button operation.
3 CCD adjustment mode <6. SPC) (*1)
An RGB adjustment mode sequence is performed (LED lights in red during CCD adjustment) when you
press the additional Reprint button during button operation.
4 Program version display & Internal Information write. <1.SPC>
The major version of a program is displayed when you press the “-” button once.
The minor version of a program is displayed when you press the “-” button second.
Indicates an error occurring in the test mode. (Lights : Error)
Progress number in the test mode
Indicates the whole test mode. (Lights : Test mode)
A signal of a test mode.