Disperse:
This function is to minimize the πCode MicroDiscs aggregations before decoding with
the PlexBio 100 Fluorescent Analyzer.
Pause:
This function pauses a program to allow opening the front door without triggering the
alarm. The ability to open the door without terminating the run provides a chance to
check microplate status, add additional reagent manually or refill/ change the solution in
the V Tray (we recommend using a new plastic V Tray for each solution).
Once the pause function is reached in an active program, the run will be stopped
temporarily. It is only safe to open the door once instructed through a pop-up window in
the User Interface. Before continuing the run, ensure that the plate is placed back onto
the heating plate, the V Tray is inserted and filled, and the door is closed. The program
will continue to the next process after the user clicks the confirm button.
Fig. 18. Disperse and Pause Procedure