3.3 System Technical Specification
Model
IW100
PERFORMANCE
Supported plates
96-well plate
(Plexbio; Cat. No. 80025 or Greiner Bio-one; Cat. No. 655101)
Processing time (full plate)
3 ~ 6.5 min (one-time wash)
Shaking
Timing setting range: 1 sec ~ 23hr 59min 59sec
Max. rate: 1200rpm
Incubator
Temperature range from RT+5°C to 60°C
Safety
Plate position sensor and door sensor
Temperature accuracy
± 0.5°C
Temperature uniformity
1°C
Shaking accuracy
≤ 10%
Dispensing accuracy
≤ ± 6%
(Measured under following conditions: 8-way manifold, 150
μL, PlexBio Wash Buffer)
Dispensing uniformity
≤ 4% CV
(Measured under following conditions: 8-way manifold, 150
μL, PlexBio Wash Buffer)
MicroDisc loss
Around ≥ 75% of MicroDiscs retained after whole processes.
(Measured under following condition: 500 ~ 1500 discs/well in
150 μL PlexBio Wash Buffer)
Residual volume
≤ 5 μl/well
(Measured under following conditions: 8-way manifold, 150
μL, PlexBio Wash Buffer)
OPERATION CONDITIONS
Temperature/Humidity
18°C
~ 32°C (64°F ~ 90°F)/20 ~ 80% RH, non-condensing
Altitude
Up to 2000 meters (6561 ft) above mean sea level
STORAGE CONDITIONS
Temperature/Humidity
-10°C
~ 70°C (14°F ~ 158°F)/10 ~ 80% RH non-condensing
PHYSICAL CHARACTERISTICS
Power
100-240V, 50-60Hz, 250W
Dimensions
Without Bottle Carrier: (W)408mm*(D)540mm*(H)470mm
With Bottle Carrier: (W)554mm*(D)540mm*(H)470mm
Weight
38 kg (84 lbs)
Fuse rating
T 5A H 250V
OTHERS
Pollution degree
2
Method of disposal
Electronic waste
Table 2. System Technical Specification