Chapter 4.
Receiving the Instrument
Note that IntelliPlex 1000 πCode Processor must be installed and serviced by PlexBio’s qualified
service engineers only. Before installation, please make sure the environmental conditions meet
the operation requirements:
Indoor use only
Operating temperature: 18°C to 32°C (64°F ~ 90°F)
Operating relative humidity: 20% to 80%, non-condensing
Operating altitude: up to 2000m (6561 ft.) above mean sea level
Storage temperature: -10°C to 70°C (14°F ~ 158°F)
IntelliPlex 1000 πCode Processor needs an operation area of 0.36 square meters and needs
to be placed at least 10 centimeters from the wall for components connection and efficient
cooling.
4.1 Inspect the Instrument Package
IntelliPlex 1000 πCode Processor arrives in a large, corrugated cardboard overpack on skids.
Inspect the shipping box and packaging for signs of damage. In case of damage, please contact
your local distributor or PlexBio Technical Support at +886-2-2627-5878 or
Do not open the box. Please schedule an installation appointment with your local distributor or
PlexBio Technical Support.
The package of IntelliPlex 1000 πCode Processor may be too heavy to
lift. Please lift using adequate equipment. Please handle with care.
Cautions
Fig. 2. The IntelliPlex 1000 πCode Processor package