Chapter 5.
Operation Procedure
Only operate the IntelliPlex 1000 πCode Processor after installation. Do not operate IntelliPlex
1000 πCode Processor if annual maintenance is overdue.
Ensure all daily/weekly/monthly maintenance is performed (refer to Chapter 7. Maintenance
for details). Not performing required maintenance will void the warranty. Moving the IntelliPlex
1000 πCode Processor is not allowed. If moving is required, please contact Customer Service of
PlexBio Co., Ltd. or qualified local distributors.
Before operation, connect the power cord and the corresponding bottles and tubes (Fig. 4).
Visually confirm that the U Tray is placed in washing tank (Fig. 5, please also refer to
section 6.1.1).
Make sure the bottles contain enough ddH
2
O and enough wash buffer (refer to
section 5.5.2), and
the waste bottle is empty.
Fig. 4. Connected diagrams
Fig. 5. Tray position
Front: SA-PE Tank (V Tray)
Back: Washing Tank (U Tray)