When used properly, image plates can be exposed, read and erased
several hundred times provided there is no mechanical damage.
•
Replace the image plate if there are any signs of damage, e.g. if the
protective layer is damaged or there are visible scratches that impair the
quality of the diagnosis.
•
Also replace the image plate if the RFID tag is damaged or becoming
detached.
•
Clean image plates properly (see section "Cleaning and disinfection" on
4.2 Operation
CAUTION
The image data on the image plate is not permanent.
The image data is altered by light, natural X-ray radiation and scattered X-
ray radiation. This will lead to a reduction in diagnostic information and
clarity.
Read the image data within 30 minutes of exposure.
Never handle exposed image plates without the light protection cover.
Do not subject an exposed image plate to X-ray radiation before or after the
scanning process.
Do not operate X-ray unit during the scanning process if the unit is in the
same room as the X-ray tube.
4.2.1 Changing input unit cartridge
The device can be used to scan image plates size 0, size 1 and size 2. Each
size of image plate requires the matching cartridge.
The size of the image plate is marked on the cartridge.
CAUTION
Loss of image information and equipment damage if an incorrect cartridge is
used.
Always use the correct size of cartridge for the image plate being used.
Before each scanning process, compare the image plate size with the LED
display on the control element.
4 Usage
User's manual
Planmeca ProScanner 2.0 35