X-ray image only shadowy
Possible cause
Remedy
The X-ray dose on the image plate was
insufficient
Increase X-ray dose.
Ghosting or double exposure on X-ray image
Possible cause
Remedy
Image plate exposed twice
Only expose the image plate once.
Image plate not sufficiently erased
Check the erasure unit is working correctly.
Inform a service technician if the problem
persists.
5 Troubleshooting
User's manual
Planmeca ProScanner 2.0 53