NOTE
The use of unsuitable agents and methods can damage the unit and
accessories.
Do not use any products based on phenolic compounds, halogen-releasing
compounds, strong organic acids or oxygen-releasing compounds, as they
may damage the materials.
•
The manufacturer recommends using disinfectants from the Dürr Dental
product range; only the products specified in these instructions have
been subjected to material compatibility testing by the manufacturer
•
Read the operating instructions for the disinfectants
Wear protective gloves.
4.3.1 Image plate scanner
Unit surfaces
The unit surface must be cleaned and disinfected of any contamination or
visible soiling.
The manufacturer recommends using the disinfectants FD 322, FD 333, FD
350 and FD 366 sensitive.
NOTE
Liquid can cause damage to the unit.
Do not spray the unit with cleaning and disinfectant agents.
Make sure that liquid does not get inside the unit.
•
Remove any soiling with a soft, lint-free cloth that has been dampened
with cold tap water
•
Disinfect the surfaces using a disinfection wipe; alternatively, use
disinfectant on a soft, lint-free cloth
Cartridges (S0-S2)
The cartridges can be cleaned and disinfected with a wipe disinfection or in
an immersion disinfection system.
NOTE
Heat damages the cartridges.
Do not thermally disinfect or steam sterilise the cartridges.
•
Remove any soiling from the cartridge with a soft, lint-free cloth that has
been dampened with cold tap water
•
Disinfect the cartridge using a disinfection wipe; alternatively, use
disinfectant on a soft, lint-free cloth
Comply with the operating instructions for the disinfectant when doing
this.
•
The manufacturer recommends ID 213, ID 212 or ID 212 forte instrument
disinfectants for the cartridges if an immersion disinfection system is
being used; these disinfectants have been subjected to material
compatibility testing
•
Allow the cartridges to completely dry before using them
4 Usage
User's manual
Planmeca ProScanner 2.0 49