No damages in the bearing surface
•
The milling unit and its zero-point must be set correctly
•
Always clamp the adapter in the same position
•
Screw connections
•
The 6 screws should always be tightened in the same order for a
reproducible positioning
•
Only screw the 6 countersunk screws loosely at first
•
Afterwards tighten the screws crosswise (always in the same
order)
7.2.2.2 Preparing adapter for test milling
NOTE
Carefully follow the directions on general cleanliness and handling.
The two test pins, available from your local premilled blank supplier, are
thoroughly inserted into the middle positions (2 + 5), pushed in to the limit,
and firmly clamped with the clamping screw by using a torque wrench.
Check that both test pins are attached directly without a gap to the
supporting surface.
7 Adapters
38 Planmeca PlanMill 50 S
User's manual