64
SSM-ICON
EN
Use, maintenance and warning manual
Static SmaRt mEtER
|
maiNtENaNcE aND FUNctiONaL cHEcKS
|
REV. B
Proceed as described in tab.9.52. (referring to Fig.9.8.) to
replace the communication battery pack
:
Step Action
1
Use the suitable tool to remove the 4 screw cover seals (
A
).
2
Unscrew the 4 fixing screws (
B
) of the transparent cover (
C
) with the Phillips screwdriver and remove it from its seat.
3
Open the door (
D
) that allows access to the communication battery compartment (
E
).
4
Disconnect the communication battery connector from its housing (
F
) and remove the battery from the com-
partment.
ATTENTION!
Store the replaced communication battery in ADR compliant packaging.
5
insert the connector of the new communication battery into its housing (
F
), then close the door (
D
) of the bat-
tery compartment making sure it fits in properly.
NOTICE!
• The connector of the battery pack is polarised so that it can only be inserted into the connec-
tor of the equipment corresponding to the correct polarity.
• Position the cables (G) so as to facilitate closing the door (D) without damaging them.
6
insert the transparent cover (
C
) in its seat, checking that it is fully inserted, then tighten the 4 fixing screws (
B
)
with the Phillips screwdriver.
NOTICE!
Tightening torque 1.2 - 1.4 Nm.
7
insert the new 4 screw cover seals (
A
) making sure they are completely inserted in the corresponding hole.
Tab. 9.52.
A
B
C
A
B
A
D
E
F
B
G
Fig. 9.8. Replacing the communication battery pack