Theory of Operation and System Architecture
Chapter 10 - Training Program
123
SpO2, and temperature measurements to digital format and to connect this
information to the processor. The respiration detection is obtained from two
of the three electrodes of the ECG connections.
Galvanic isolation of these circuits from the remainder of the monitor is
accomplished by utilizing an isolated power supply and by incorporating
opto-isolators between the Front End outputs and inputs to the microprocessor
computation and control circuitry.
NiBP Front End
The NiBP section contains the pumps, valves, pressure measurement
circuitry, and control circuitry for the noninvasive blood pressure
measurement. Pressure data is converted to digital format and conveyed to the
processor section.
Power System
The power system section contains a power supply capable of operating the
monitor and charging the battery from either an AC source of 100 to 240 volts
AC at 50 to 60 Hz or a DC input of 10 to 16 volts. This section also contains
the circuitry. The battery provides the operating power for the monitor.
Power system outputs of ±12 volts DC, +5 volts DC, and –24 volts DC are
developed for use throughout the non-isolated portions of the monitor.
Microprocessor,
Memory and
Control
The microprocessor (µP), Memory and Control section contains the system
CPU and all digital support circuitry. The latter includes the RAM,
nonvolatile memory, and real-time clock. This section also contains the
display logic, keypad (switch) interface logic, RS-232 I/O control,
defibrillator synchronization control, and printer logic.
Display
The display is a cold-cathode, backlit, fluorescent LCD unit. The pixel
resolution of 0.23 mm provides a 640 x 480 line display. For the A3 models
M3926A and M3927A, the display is presented in black and white mode only
(no gray scale). The A3 models M3928A and M3929A have color a display.
Keypad (Switches)
and Wheel
The keypad circuit contains five push-switch membrane switches and two
green LEDs. The LEDs are driven by the power supply system and indicate
(AC or DC), the charging of the batteries, and the condition of the source
cable in-line fuses.
Signals from the LEDs are returned to the microcomputer for processing and
control as required.
The power switch, connected directly to the supply, toggles the power
between Standby and On modes. When in Standby, the display is blank, and
Summary of Contents for M3921A
Page 12: ...xii Contents ...
Page 16: ...Related Documents 4 Introduction ...
Page 48: ...Safety Tests 36 Performance Verification ...
Page 72: ...Troubleshooting Guide 60 Chapter 5 Troubleshooting ...
Page 100: ...Disassembly Procedures A3 88 Chapter 6 Disassembly Guide ...
Page 103: ...Top Level Assembly Chapter 7 Spare Parts 91 Figure 2 A1 Top Level Assembly Drawing 2 of 2 ...
Page 105: ...Top Level Assembly Chapter 7 Spare Parts 93 Figure 3 A3 Top Level Assembly Drawing ...
Page 108: ...Front Case Assembly 96 Chapter 7 Spare Parts Figure 5 A1 Front Case Assembly Drawing 2 of 3 ...
Page 110: ...Front Case Assembly 98 Chapter 7 Spare Parts Figure 7 A3 Front Case Assembly Drawing 1 of 2 ...
Page 111: ...Front Case Assembly Chapter 7 Spare Parts 99 Figure 8 A3 Front Case Assembly Drawing 2 of 2 ...
Page 114: ...Rear Case Assembly 102 Chapter 7 Spare Parts Figure 9 A3 Rear Case Assembly Drawing 1 of 3 ...
Page 115: ...Rear Case Assembly Chapter 7 Spare Parts 103 Figure 10 A3 Rear Case Assembly Drawing 2 of 3 ...
Page 116: ...Rear Case Assembly 104 Chapter 7 Spare Parts Figure 11 A3 Rear Case Assembly Drawing 3 of 3 ...
Page 124: ...Repacking In A Different Carton 112 Chapter 8 Packing For Shipment ...
Page 144: ...Preventative Maintenance 132 Chapter 10 Training Program ...
Page 150: ......