Diagnostic Software
EN 40
LX7500R
5.
Basic Engine (BE)
Nucleus Name
DS_PSCAN_DevTypeGetDeinterlacer
Nucleus Number
807
Description
Get the device (revision) type information of the progressive scan deinterlacer.
User Input
None
Example
DS:> 807
080700:
Chip name : 2300
Chip version : 1
Test OK @
Nucleus Name
DS_PSCAN_CommunicationDeinterlacer
Nucleus Number
808
Description
Check the communication between the IIC controller of the chrysalis and the progressive
scan Deinterlacer-IC
User Input
None
Example
DS:> 808
080800:
Test OK @
Nucleus Name
DS_BE_CommunicationEcho
Nucleus Number
900
Description
Check the communication between the digital board and the basic engine by issuing an
echo command over the S2B interface
User Input
None
Example
DS:> 900
090000:
Test OK @
Nucleus Name
DS_BE_Reset
Nucleus Number
901
Description
Reset the basic engine
User Input
None
Example
DS:> 901
090100:
Test OK @
Nucleus Name
DS_BE_VersionGet
Nucleus Number
903
Description
Get the version of the basic engine and that of the optical unit
User Input
None
Example
DS:> 903
090300: BE version = 20.09.18 Optical unit version = 3C.00.09.41.08
Test OK @
Nucleus Name
DS_BE_GetSelftestResult
Nucleus Number
902
Description
Return the self-test results through the service port
User Input
None
Example
DS:> 902
090200:
Test OK @
Nucleus Name
DS_BE_TrayOut
Nucleus Number
904
Description
Open the tray of the basic engine
User Input
None
Example
DS:> 904
090400:
Test OK @
Nucleus Name
DS_BE_TrayIn
Nucleus Number
905
Description
Close the tray of the basic engine