15

3.12. Connecting The Serial 

Interface RS 485

An external operating component DCU 001/DCU 600 o

orr

HPU 001 o

orr – using a level converter – an external computer

can be connected via the connection ”RS 485” on the TC 750
with the use of a shielded 8 pole modular connecting cable
contained with the delivery.

PIN

Arrangement

1

not connected

2

+24V output (

210 mA loading 

capacity)

3

not connected

4

not connected

5

RS 485: D+ (DO / RI)

6

Gnd

7

RS 485: D- (DO / RI)

8

not connected

RS 485

1 ... 8

   

      

 

RT

 

 

 

RT

 

  

Level Converter

 

    

  

 

  

 

TC

 

PC

TC

5:D+, 7:D-

The electrical connections are internally optically decoupled.

Description

Value

Serial Interface Type:

RS 485

Baud rate:

9600 baud

Data file word length:

8 bit

Parity:

no (no parity)

Start bits:

1

Stop bits:

1..2

(View from the
plug side of the
TC 750)

It is possible to connect an RS 232 (e.g. PC) via

a level converter (please see Section 11.
Accessories).

PLEASE NOTE

All units connected to the bus must have differing serial 
interface addresses (parameter 797).
The group address of the TC 750 is 960.

Only safety low voltages (SELV) may be 

connected to Serial Interface RS 485 .

All switched on remote functions have priority over the serial
interface functions.

Please refer to the separate Operating Instruc-

tions PM 0488 BN for more detailed 
information on operations via Serial Interface
RS 485 and its electronic data.

Profibus DP Gateway TIC 250 is available (see Section 11.
Accessories) for connecting an Electronic Drive Unit TC 750
to a Profibus DP.

Please refer to the respective Operating Instructions 
PM 0599 BN for detailed information on the operation of the
TIC 250.

PLEASE NOTE

WARNING

Connecting The RS 485

Connecting to a fixed bus system:

Connect all units with D+ (pin 5/RS 485) and 
D- (pin7/RS 485) to the bus.

The bus must be completed at both ends.

The connections should be made in accordance with the
specification of the Serial Interface RS 485.

The serial interface is galvanically and safely separated from
the maximum supply voltage from the TC 750.

Connection

 

 

DCU 001

without power supply

 

 

       

 

   

 

      

    

   

 

   

 

PC

external 
computer

DCU 600

with power supply

or

or

TC 750

HPU 001

RS 485

RS 232

level converter

RS 485

RS 485

RS 485

RS 485

Connecting the Serial Interface RS 485

Summary of Contents for TPH 1201 P

Page 1: ...Turbomolecular Pump With Electronic Drive Unit TC 750 PT 0147 BE G 0703 TPH 1201 P PC TPU 1201 P PC Betriebsanleitung Operating Instructions...

Page 2: ...erations 18 Operation with cooling water temperature 25 C 19 4 5 Circulatory Lubrication 20 4 6 Switching OFF And Venting 20 4 7 Shutting Down For Longer Periods 20 4 8 Operations With The DCU 001 DCU...

Page 3: ...the turbopump Take care to avoid burns Keep leads and cables well away from hot surfaces 70 C Operate the turbopump with TC 750 only in conjunction with the relevant power supply accessory Without th...

Page 4: ...y II Connection voltage 140 VDC 5 Proper Use The Turbomolecular Pumps may only be used for the purpose of generating vacuum The Turbomolecular Pumps TPH TPU 1201 P C please see the type plate are desi...

Page 5: ...been agreed by the manufacturer The connection to power supplies with earthing of a direct voltage pole The turbomolecular pumps and flange connections must not be used to tread on when climbing onto...

Page 6: ...ccidental contact Floor mounting of the turbomolecular pump is admissible after consulting the manufacturer The person responsible for commissioning must ensure that the installation is carried out in...

Page 7: ...he component parts in your set of mounting material Use 1 18 8 bracket screws Tighten the clamps crosswise in three steps Tightening torque 5 Nm 15 Nm 25 2 Nm See that the sealing surface is not damag...

Page 8: ...your set of mounting material Use 1 12 2 screws and 2 24 4 nuts For a flange material having a strength of 170 N mm2 and a stud screw with clearance hole the nuts must be tightened crosswise in three...

Page 9: ...Pfeiffer Vacuum range Connecting the backing pump All connections of the fore vacuum line With the usual small flange components or screwed hoses Be sure to conduct away the exhaust gases from the bac...

Page 10: ...nnection to a torque of 10 Nm Tighten all hose clips and check for firm seating of the hoses Cooling with the Recycled Water Cooling Unit TZK Accessory Connection Dirt traps are not permissible in the...

Page 11: ...TC 750 should be disconnected and then proceed in accordance with the relevant Operating Instructions 1 only with cable PM 051 843 T please see Section 11 CAUTION PLEASENOTE 3 7 Installing The Power S...

Page 12: ...serts with screwed cable glands and the cover 107d and insert the cables at the TC 750 In case the pump is operated using the remote control the remote control unit must be connected in accordance wit...

Page 13: ...agonal SW 14 Electrical connection Plug control lead 42a into the connection VENT of the TC 750 8 on the turbopump Release for venting is executed via remote plug pin 2 pre set on delivery please see...

Page 14: ...be processed 6 standby off standby on pump is accelerated to 66 of its nominal rotation speed 7 rotation speed setting mode off the rotation speed can be changed by feeding a PWM signal to this pin o...

Page 15: ...the TC 750 is 960 Only safety low voltages SELV may be connected to Serial Interface RS 485 All switched on remote functions have priority over the serial interface functions Please refer to the separ...

Page 16: ...Mains input heating 115 208 230 VAC HEATING L N PE PE N L REMOTE 1 2 3 4 5 6 24 VDC max 50 mA Venting release Motor TMP Pumping station Heating Reset opt sealing gas Standby PWM on touch relation 25 7...

Page 17: ...lay And Operating Unit DCU HPU Turbopump rotors rotate at high speed When the high vacuum flange is open there is a danger of injury and of damage to the pump caused by objects falling into the pump T...

Page 18: ...a the DCU or HPU please refer to the respective operating Instructions Pumping gases with molecular mass 40 with the incorrect gas mode can cause damage to the pump When pumping noble gases heavier th...

Page 19: ...to high for the current power the rotor can overheat and being damaged that way A cooling water temperature 25 C has no influen ce the pump acts as with a cooling water temperature of 25 C To reduce p...

Page 20: ...ble First stage venting with a pressure increase rate of 15 mbar s for 20 seconds Second stage venting with an optionally large venting valve The valve cross section for a venting rate of 15 mbar s mu...

Page 21: ...ed setting operations for the turbopump Switching Outputs Switching outputs 1 and 2 can be loaded with a maximum 24 V 50 mA per output The following functions are assigned to the switching outputs S S...

Page 22: ...ts in the turbomolecular pump being switched off LED on the TC 750 glows red Collective malfunction L LE ED D C Ca au us se e g gr re ee en n r re ed d glows Power supply OK Function pumping station O...

Page 23: ...CU HPU or PC Thermal overloading caused by Insufficient water flow Ensure free flow Insufficient air supply Ensure adequate air supply Fore vacuum pressure too high Reduce fore vacuum pressure Ambient...

Page 24: ...FR 49 Part 173 240 Definition and Preparation We will carry out the decontamination and invoice this work to you if you have not attached this note This also applies where the operator does not have t...

Page 25: ...tle the turbopump to change the lubricant this may only be carried out in the position shown in Section 3 2 otherwise the lubricant could contaminate the turbopump It is possible to detach the lubrica...

Page 26: ...Section 11 1 Install the lubricant pump back onto the turbomolecular pump again 8 4 Replacing The Electronic Drive Unit TC 750 Turbopump and Electronic Drive Unit TC 750 must only be disconnected fro...

Page 27: ...ow rate for Nitrogen N2 l s 1250 1250 1170 1170 Helium He l s 1300 1300 1230 1230 Hydrogen H2 l s 1100 1100 1050 1050 Argon Ar l s 1200 1200 1100 1100 CF4 l s 950 950 890 890 Compression ratio N2 1 10...

Page 28: ...sions Diagram 240 200 142 9 40 70 314 242 5 191 123 122 334 5 25 M16x32 tief 4 0 3 0 1 2 0 96 4 5 73 264 69 95 140 4 161 128 17 30 30 30 99 170 0 1 M6 DN 200 ISO K DN 40 ISO KF M6 G1 8 TPH 1201 P P C...

Page 29: ...29 200 142 9 40 70 122 25 M16x32 4 0 3 0 1 2 0 96 4 5 73 69 95 140 161 128 17 30 30 30 99 264 341 5 130 197 91 249 5 321 170 0 1 285 M6 DN 40 ISO KF DN 200 ISO F M6 G1 8 TPH 1201 P P C DN 200 ISO F...

Page 30: ...30 200 142 9 40 70 122 25 M16x32tief 4 0 3 0 1 2 0 96 4 5 73 69 95 140 161 99 253 264 142 210 262 333 354 128 30 30 30 17 170 0 1 M6 DN 200 CF F DN 40 ISO KF M6 G1 8 TPU 1201 P P C DN 200 CF F...

Page 31: ...rive Unit TC 750 1 PM C01 713 8d Remote plug 1 P 0920 668 E 17 Seal ring for water connections 2 P 3529 133 A 71 Lubricant filler screw 1 N 3191 382 X see Section 4 72 Lubricant drain screw 1 N 3191 3...

Page 32: ...321 Other accessories Connecting cable TC 750 TPS DCU PM 051 103 T 3 m other lengths on request Connecting cable TC 750 TPS DCU shielded PM 051 143 GT 3 m other lengths on request Power Supply TPS 60...

Page 33: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 34: ...specified below is intended for installation in a machine which is forbidden to be put into operation until such time as it has been determined that the end product is in accordance with the provisio...

Page 35: ...0 64 41 8 02 2 02 info pfeiffer vacuum de www pfeiffer vacuum net Vacuum is nothing but everything to us Turbopumps Rotary vane pumps Roots pumps Dry compressing pumps Leak detectors Valves Component...

Reviews: