Drive motors
Drive motors
Removal
1. Raise the seat to the highest position. If the
wheelchair is equipped with a fixed seat
post or if the seat lift does not work normally
because the batteries are discharged or the
adjustment device is defective, the seat can be
raised/lowered manually.. See page 29.
2.
Switch off the main power switch on the
control panel.
3.
Remove the chassis covers. See page 7-8.
4.
Lift up and chock up the wheelchair chassis so
that the wheel in question is free of the ground.
5.
Remove the drive wheel. See page 16.
6.
Separate the magnetic brake and drive motor
cabling at the contacts on the cabling. These
are positioned on the inside of the chassis,
next to the seat lift. (see fig.). The contacts are
mounted together in a holder.
7.
Run the connection cables out through the
chassis cable duct.
8.
Remove the drive motor, it’s fitted with four
screws (see fig.).
The contacts are mounted together in a holder.
For this task the following tools are necessary:
1 Allen key 5 mm.
1 Allen key 6 mm.
Fitting/removing the Drive motor.
1
Screw ISO 4762 M6x60 8.8 Fe/Zn
2
Washer, ISO 7089 6 200 HV Fe/Zn 5 C1(6,4x12x1,6)
Item
Description
2
1
Summary of Contents for M300 Corpus HD
Page 1: ...SERVICE MANUAL Permobil M300 Corpus HD Power Wheelchair US ...
Page 2: ......
Page 103: ...103 Troubleshooting ...
Page 104: ...104 Cabling overview A B C D E F 1 2 3 4 ...
Page 105: ...105 Cabling overview Cabling overview Chassis A B C D E F 5 6 7 8 ...
Page 106: ...106 Cabling overview A B C D E F 1 2 3 4 ...
Page 107: ...107 Cabling overview Cabling overview Seat A B C D E F 5 6 7 8 ...
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ...Article no 205270 US 0 ...