Tools menu
99
Filter properties
The filters used in the instrument have the properties shown in the table below.
The following abbreviations are used:
CWL = Centre wavelength, BW = Bandwidth (Full Width Half Maximum), Tpk =
Maximum transmittance at the peak wavelength (%)
Filter type
Name
Centre
Wavelength
(CWL)
Bandwidth
(BW) (Full
Width Half
Maximum)
Maximum transmittance
at peak wavelength (Tpk)
(%)
Tr-fluorescence excitation
D340
340 nm,
ca 50 nm
ca 60%
Tr-fluorescence excitation
D320
320 nm,
ca 80 nm
ca 70%
Europium chelate emission
(LANCE option installed)
D615
615 nm,
ca 8.5 nm
ca 80%
Europium chelate emission
(LANCE option installed)
D665
665 nm
ca 7.5 nm
ca 80%
Samarium chelate emission
D642
642 nm,
ca 6.5 nm
ca 80%
Terbium chelate emission
D545
545 nm,
ca 7.5 nm
ca 70%
Dysprosium chelate emission
D572
572 nm,
ca 7.5 nm
ca 70%
Colorimetric label emission
P405
405 nm,
ca 10 nm
ca 45%
Colorimetric label emission
P450
450 nm,
ca 6 nm
ca 30%
Colorimetric label emission
P490
490 nm,
ca 6 nm
ca 30%
Fluorescein excitation
F485
485 nm
ca 15 nm
ca 60%
Fluorescein emission
F535
535 nm
ca 25 nm
ca 50%
Umbelliferone excitation
F355
355 nm
ca 40 nm
ca 35%
Umbelliferone emission
F460
460 nm
ca 25 nm
ca 50%
Note: the time-resolved filters are available only if this option is included with the
instrument.
Summary of Contents for Wallac 1420
Page 2: ......
Page 10: ......
Page 11: ...9 Introduction...
Page 12: ...10...
Page 17: ...15 Operating Wallac 1420...
Page 18: ...16...
Page 36: ...Operating Wallac 1420 34...
Page 37: ...35 Protocol editing...
Page 38: ...36...
Page 59: ...57 Result viewing...
Page 60: ...58...
Page 67: ...65 Tools menu...
Page 68: ...66...
Page 128: ......
Page 129: ...127 Exiting...
Page 130: ...128...
Page 132: ...Exiting 130...
Page 133: ...131 Troubleshooting...
Page 134: ...132...
Page 138: ...Troubleshooting 136...
Page 139: ...137 Appendices...
Page 140: ...138...
Page 165: ...163 Index...
Page 166: ...164...