Using the RamanMicro 200 . 83
General SOPs
The RamanMicro 200 combined with the Spectrum software is a very flexible instrument.
This chapter demonstrates how to perform a number of tasks. The procedures provided may
not be the most straightforward or simple for the task you wish to perform, so treat these
Standard Operating Procedures (SOPs) as a starting point and with time you may wish to
develop your own SOPs which better suit your specific mode of use. These procedures also
act as a “learn by doing” section allowing you to explore the functionality of the software and
spectrometer.
The General SOPs demonstrate how to collect data when your samples are on the
microscope stage. For SOPs that demonstrate how to collect data using a fiber optic probe,
please refer to
Fiber Optic Probe SOPs
on page 92. For step-by-step instructions for
calibrating the spectrometer, see
Calibration and Performance Verification SOPs
on page 94.
SOP-1: Turning the Instrument On
1.
Switch on the PC and monitor.
2.
Login to Windows.
3.
Ensure the laser power key is in the ARM position on the front panel of the
spectrometer.
4.
Ensure the power supply is connected to a mains outlet and that the outlet is switched
on. If you have a motorized stage, you will need to switch on the power supply unit.
5.
To turn on the spectrometer, press the PSU power switch ON (Figure 30):
- ON
Ο
- OFF
Figure 30 The PSU power switch
Summary of Contents for Raman Micro 200 Series
Page 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanMicro 200 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 29: ...Raman History Theory and Instrumentation...
Page 35: ...Preparing Your Facility...
Page 41: ...Installing Your RamanMicro 200...
Page 76: ...76 RamanMicro 200 Series Getting Started Guide...
Page 77: ...Using the RamanMicro 200...
Page 103: ...Maintenance and Troubleshooting...