100 .
RamanMicro 200 Series Getting Started Guide
11.
Repeat steps 2–10 for each objective lens.
12.
Remove the intensity correction standard and store safely.
Any data acquired will now be intensity corrected.
Remember that if you have a fiber optic probe, you must intensity calibrate the fiber optic
probe as well as the microscope objective lenses. Ensure that the fiber optic probe has a lens
cap in place during intensity calibration of the objectives.
Summary
•
The only calibration procedure that needs to be performed on a regular basis is the
wavelength calibration verification procedure.
•
The wavelength and laser frequency calibrations should only be performed if the
instrument fails the wavelength calibration verification procedure.
•
The intensity calibration should only be performed on a yearly basis or if the
environment (temperature and / or humidity) changes significantly, or if the instrument
is moved or repaired.
Summary of Contents for Raman Micro 200 Series
Page 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanMicro 200 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 29: ...Raman History Theory and Instrumentation...
Page 35: ...Preparing Your Facility...
Page 41: ...Installing Your RamanMicro 200...
Page 76: ...76 RamanMicro 200 Series Getting Started Guide...
Page 77: ...Using the RamanMicro 200...
Page 103: ...Maintenance and Troubleshooting...