![PerkinElmer Raman Micro 200 Series Getting Started Manual Download Page 44](http://html1.mh-extra.com/html/perkinelmer/raman-micro-200-series/raman-micro-200-series_getting-started-manual_1530231044.webp)
44 .
RamanMicro 200 Series Getting Started Guide
Attaching the External Interlock Override
The RamanMicro 200 Series has an external interlock connection located on the rear panel of
the spectrometer module. To operate the instrument, the external interlock must be
satisfied.
You must connect an external door interlock to the 7-way DIN connector labeled
INTERLOCK on the rear panel of the spectrometer, or attach the External Interlock Override
(L1320264) supplied with the instrument.
If your laser safety officer determines that an interlock is required on the door to the room,
you will need to purchase the Raman Laser Safety Kit (L1320262). For more information
contact your PerkinElmer representative.
Summary of Contents for Raman Micro 200 Series
Page 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanMicro 200 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 29: ...Raman History Theory and Instrumentation...
Page 35: ...Preparing Your Facility...
Page 41: ...Installing Your RamanMicro 200...
Page 76: ...76 RamanMicro 200 Series Getting Started Guide...
Page 77: ...Using the RamanMicro 200...
Page 103: ...Maintenance and Troubleshooting...