![PerkinElmer Raman Micro 200 Series Getting Started Manual Download Page 46](http://html1.mh-extra.com/html/perkinelmer/raman-micro-200-series/raman-micro-200-series_getting-started-manual_1530231046.webp)
46 .
RamanMicro 200 Series Getting Started Guide
Connecting the Motorized Stage
The motorized stage is powered by the stage controller box (L1860289) and joystick
(L1865723).
Figure 15 Stage controller and joystick
To connect the motorized stage:
1.
Connect the cable from the joystick to the socket labeled JOYSTICK on the rear panel
of the stage controller box (Figure 16).
Figure 16 Rear panel of the Stage Controller
2.
Connect the cable from the socket on the right side of the stage to the MOTOR X, and
MOTOR Y sockets on the rear panel of the stage controller box.
3.
Connect the RS232 cable (L1865749) from the PC to the socket labeled RS232 on the
rear panel of the stage controller.
4.
Connect the power cable from the stage controller box to the mains supply.