Problem:
During system initialization the laser does not stabilize.
Description The laser initialization form displayed during system initialization runs for
over 60 seconds. During this time the laser should have stabilized.
Solution
Ensure that the laser key on the front panel of the spectrometer is set to the
ARM position.
Problem:
Detector cannot achieve operational temperature.
Description During system initialization a message box stating, ‘Detector temperature
could not be achieved’ is displayed.
Solution
Close and open the software.
Problem:
Need to upgrade to a new version of Spectrum software.
Description You must uninstall the current version of Spectrum before installing a new
version.
Solution
To uninstall Spectrum software, open the Control Panel and then select
Add or Remove Programs. Then remove the Philips ToUcom Pro
Camera, the Raman Instrument and then Spectrum Software (in that
order). You should also delete files C:\pel_apps\, C:\pel_data, C:\Documents
and Settings\All Users\Application Data\PerkinElmer\SecuritySystem,
C:\Windows\pel_apps.ini and C:\Windows\pe_sopb.ini using Windows
Explorer.
To install the new version of Spectrum software follow the instructions in
Installing Spectrum Software
on page 54.
Summary of Contents for Raman Micro 200 Series
Page 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanMicro 200 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 29: ...Raman History Theory and Instrumentation...
Page 35: ...Preparing Your Facility...
Page 41: ...Installing Your RamanMicro 200...
Page 76: ...76 RamanMicro 200 Series Getting Started Guide...
Page 77: ...Using the RamanMicro 200...
Page 103: ...Maintenance and Troubleshooting...