Figure 25 Four retaining screws in the sample compartment
2.
Lift up the sample compartment off the locating pins and remove it from the
spectrometer.
Removing the Detector Unit
WARNING
Electrical Hazard
The connector for the detector unit carries high voltage (130 V dc).
Switch off and disconnect the instrument before removing or fitting a
detector/accessory unit.
1.
Switch off the spectrometer and disconnect the line power cord.
2.
Open the sample compartment cover.
3.
Undo the two retaining screws located under the blanking caps on the top of the
detector (Figure 26) and are accessible using the long posidrive screwdriver supplied
with the instrument.
Figure 26 Undoing the retaining screws
Retaining Screw
Retaining Screw
Retaining Screws