58 . Maintenance
Figure 17 Drain holes in standard sample compartment baseplate
1.
First remove the cell holder or other sample handling accessory from the sample
compartment.
2.
Using a soft cloth and mild laboratory detergent solution, lightly scrub away all foreign
material.
3.
Using a clean cloth dampened with water, rinse the cleaned surfaces thoroughly.
4.
Dry with lint free cloth or tissue.
Sample Compartment Windows
Four windows are provided with the spectrometer. The windows are made of silica and may
be used in the entire spectral range of the spectrometer.
The windows seal the sample compartment and thus protect the instrument’s optics from
dust and fuming or aggressive samples.
•
Generally, the windows should be installed at all times.
•
The windows are an optical component and require the same care and handling as cells.
•
The windows can be removed to clean them. Each window has a magnetic frame and
can be carefully removed by hand.
Windows are most suitably cleaned by wiping them with a soft cloth moistened with
ethanol.
Replace the window in its original position after cleaning.
Use and Care of Cells
A good spectrometer cell is an optical device, forming a part of the optical system of the
instrument with which it is used. It must be accorded the same careful treatment applied to
Drain Holes
Summary of Contents for Lambda 650
Page 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Introduction ...
Page 11: ...Safety Practices ...
Page 23: ...System Description ...
Page 32: ...32 System Description ...
Page 33: ...Lambda 650 Installation ...
Page 44: ...44 Lambda 650 Installation ...
Page 45: ...Using the Instrument ...
Page 54: ...54 Using the Instrument ...
Page 55: ...Maintenance ...
Page 72: ...72 Maintenance ...