Lambda 650/850/950 Hardware Guide . 43
•
Manual set 2 peaks
enables you to change the offset and factor by specifying
the old and new values of 2 well known peaks by entering values of
Old Peak 1
,
New Peak 1
,
Old Peak 2
and
New Peak 2
.
•
Auto search 1 peak
enables you to change the offset. The software performs an
automatic search for the D
2
peak at 656.1 nm. The measured peak is then shifted
to the exact wavelength.
•
Auto search 2 peaks
enables you to change the offset and factor. The software
performs an automatic search for the D
2
peak at 656.1 nm and the peak at 0.0 nm.
The measured peaks are then shifted to the exact wavelengths.
0%T
This calibrates the electronic offsets.
¾
Select
Auto
to enable the software to determine the required offset.
OR
Enter the required
Calibration offset
value to be used.