O
PERATING
S
OFTWARE
OS6.0
/
OS10.0
55
ANALOG START
This parameter is used to set the start value of the analog modulation.
The start value specifies the measurement result at which the analog output
modulates 0 V or 0/4 mA
.
Value
Function
-99999999
Smallest start value
0
Default value
99999999
Largest start value
ANALOG END
This parameter is used to set the end value of the analog modulation. The end
value specifies the measurement result at which the analog output modulates its
max. value (+/-) 10 V or 20 mA.
Value
Function
-99999999
Smallest end value
10000
Default value
99999999
Largest end value
ANALOG GAIN (%)
This parameter is used to set the maximum modulation. The ANALOG GAIN
indicates the maximum modulation of the analog output in % relative to (+/-) 10 V
or 20 mA.
Example 1: 102.00 corresponds to a modulation of 10.2 V / 20.4 mA, once the
ANALOG END value is reached.
Example 2: 95.00 corresponds to a modulation of 9.5 V / 18 mA, once the
ANALOG END value is reached.
Value
Function
0.00
Smallest modulation
100.00
Default value
110.00
Largest modulation
ANALOG OFFSET (%)
This parameter is used to set the zero point shift of the output.
Example: 0.20 corresponds to a modulation of 0.02 V / 0.04 mA, when the
ANALOG START value is reached.