PCE Instruments PCE-DDM 10 Manual Download Page 8

Manual 

5.2 

Wireless hand-held display unit 

   

 

 

 

 

1) “

Zero

” 

a) When the device is switched on without any load attached to the hook and the display still shows 
numbers, press this key to set the device to 0. 
b) In the settings this key is used as the up key. 
 
2) “

Tare

” 

a) Attach a loop or a container to the hook and press this key when the device displays a stable value. 
The display will now show “0” and “N” flashes. 
b) Pour the products into the container. The device will now show the net weight of the product. 
c) When you remove the products and the container from the hook, the device will display negative 
values. 
d) In the settings this key is used to switch between the digits. 
 
3) “

Mode

” 

Press “MODE” to switch the backlight off and “ENTER” to confirm this. 
Press “MODE” twice to get to the setup and “ENTER” to get to the parameter settings. Press “MODE” to 
return to measurement mode. 
Press “MODE” 3 times. The display will show “dC”. Then press the “ENTER” key to check the battery 
level. Press “MODE” to return to measurement mode. 
Press “MODE” 4 times. The display will show “Unit”. Then press “ENTER” to get to the mode in which you 
can select the unit. Press “Zero” to select the unit and “Mode” to return to measurement mode. 
Press “MODE” 5 times. The display will show “ACCU”. Then press “ENTER” to check the memory status. 
Press “MODE” to return to measurement mode. 
Press “MODE” 6 times. The display will show “SACCU”. Press “ENTER” to search within the memory. 
Press “MODE” to return to measurement mode. 
Press “MODE” 7 times. The display will show “CLEAr”. Press “ENTER” to delete the memory. Press 
“MODE” to return to measurement mode. 
Press “MODE” 8 times. The display will show “CLIBr”. Press “ENTER” and hold it for 3 seconds to check 
the calibration process. Press “ENTER” and hold I for fewer than 3 seconds to return to measurement 
mode. 
Press “MODE” 9 times. The display will show “ESC”. Press “ENTER” and “Peak” to return to the default 
settings. 
 
4) “

Enter

” 

Is used to select a function, normally in combination with the “MODE” key. 
 
5) “

Peak

” 

When you attach a new weight and press the PEAK key, the device will show the maximum value of this 
new weight and freezes it into the display. By pressing the key again, you return to normal measurement. 

Summary of Contents for PCE-DDM 10

Page 1: ...nts UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From Outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instr...

Page 2: ...ide 9 5 3 1 ON OFF 9 5 3 2 Zero 9 5 3 3 TARE 9 5 3 4 PEAK HOLD 9 5 3 5 Data memory 9 5 3 6 Data search 9 5 3 7 Delete memory 10 5 3 8 Battery voltage 10 5 3 9 To change the unit 10 5 3 10 To return to...

Page 3: ...of the measuring equipment must be borne in mind for example those of the crane producer of the load suspension device producer and of the shackle producer In this relation the measurement equipment m...

Page 4: ...ar of the attachment parts Control all dimensions Consider all possible deformations and damage Make sure that all parts work properly Every 12 months or 50000 measurements whatever happens first Insp...

Page 5: ...ies operation time approx 50 h battery life 40 hours model range resolution minimum load weight with shackle PCE DDM 3 3000 kg 30 kN 1 kg 10 N 10 kg 2 kg 5 kg PCE DDM 5 5000 kg 50 kN 2 kg 20 N 40 kg 3...

Page 6: ...mm B mm C mm D mm mm H mm Material PCE DDM 3 260 123 37 195 51 365 aluminium PCE DDM 5 285 123 57 210 58 405 aluminium PCE DDM 10 320 120 57 230 92 535 steel PCE DDM 20 375 128 74 260 127 660 steel PC...

Page 7: ...ta memory f STB stable condition g kN kilonewton N newton in taring mode the N flashes h t tonne i lb lbs k kg kilogrammes 5 1 2 Keypad a ON OFF 1 Press the ON OFF key to switch the device on 2 When t...

Page 8: ...to measurement mode Press MODE 4 times The display will show Unit Then press ENTER to get to the mode in which you can select the unit Press Zero to select the unit and Mode to return to measurement...

Page 9: ...switches off automatically after 60 minutes 5 3 2 Zero OPERATION DISPLAY When the device is switched on it normally displays In case it shows small numbers press this key 1 Press ZERO 2 Press ZERO 5 3...

Page 10: ...calibration You can use the hand held device to select parameter settings or calibration 5 4 1 Key functions Key Function Tare to change numbers ZERO to increase values MODE to save and exit ENTER to...

Page 11: ...rated it will display the calibrated value Press Zero to select Select the unit Un 0 kg Un 1 lb Un 2 N Un 3 KN Un 4 t Normally select Un 0 Press Enter to confirm If the device is new it will display O...

Page 12: ...after attaching the tare weight You can also attach the tare weight before switching the device on Press MODE 8 times The display will show CLIBr Press ENTER and hold it for 3 seconds to confirm the c...

Page 13: ...attached before STB light flashes 5 6 Troubleshooting problem possible reason solution no display batteries defective replace key defective call support ON OFF key has not been pressed properly press...

Page 14: ...a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 7 Contact If you have any questions about our range of products or measurin...

Reviews: