PCE Instruments PCE-DDM 10 Manual Download Page 10

Manual 

10 

5.3.7 

Delete memory 

OPERATION 

DISPLAY 

Description

 

Press “Mode” 7 times and 

confirm with “Enter”. 

 

question: delete or not? 

 

Press “Mode” to stop. 

 

Press “Enter” to delete. 

5.3.8 

Battery voltage 

OPERATION 

DISPLAY 

Description

 

Press “Mode” 3 times. Display 

shows “dC”. Press “Enter”. 

 

display  power  battery  voltage 

Press “Mode”. 

 

back 

5.3.9 

To change the unit 

OPERATION 

DISPLAY 

Description

 

Press “Mode”. Select “Unit” and 

press “Enter”. 

 

Press “Enter”. The units are 

changed every time “Enter” is 

pressed. 

Press “Mode” again. 

 

back 

5.3.10  To return to measurement mode 

OPERATION 

Description

 

Press “Mode”. Select “ESC” and 

press “Enter”. 

to return to measurement mode 

5.4 

Parameter selection and calibration 

You can use the hand-held device to select parameter settings or calibration. 

5.4.1 

Key functions 

Key 

Function 

“Tare” 

to change numbers 

“ZERO” 

to increase values 

“MODE” 

to save and exit 

“ENTER” 

to confirm 

5.4.2 

Parameter settings 

OPERATION 

DISPLAY 

Description

 

Press “Mode” twice. 

 

to confirm, open the parameter 

settings 

Press “Enter”. 

 

device displays capacity in case it 

is calibrated 

Press “Zero” to select. 

 

select  capacity 

The display shows  

02/03/05/06/10/15/20/30/50/60/75 

one after the other. Select cap: 

2000 kg and for example FS =02. 

 

Press “Enter” to confirm. 

 

The display shows ld = 02. If the 

device is calibrated, it will show 

the calibrated range. 

Press “Zero” to select. 

 

select division 

The display shows  

01/02/05/10/20 one after the 

other.  2000 kg 

The 

division is 1 kg, so choose ld = 

01. 

 

 

If the device is new, it will display 

Pt = 0. If the device is calibrated, 

Summary of Contents for PCE-DDM 10

Page 1: ...nts UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From Outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instr...

Page 2: ...ide 9 5 3 1 ON OFF 9 5 3 2 Zero 9 5 3 3 TARE 9 5 3 4 PEAK HOLD 9 5 3 5 Data memory 9 5 3 6 Data search 9 5 3 7 Delete memory 10 5 3 8 Battery voltage 10 5 3 9 To change the unit 10 5 3 10 To return to...

Page 3: ...of the measuring equipment must be borne in mind for example those of the crane producer of the load suspension device producer and of the shackle producer In this relation the measurement equipment m...

Page 4: ...ar of the attachment parts Control all dimensions Consider all possible deformations and damage Make sure that all parts work properly Every 12 months or 50000 measurements whatever happens first Insp...

Page 5: ...ies operation time approx 50 h battery life 40 hours model range resolution minimum load weight with shackle PCE DDM 3 3000 kg 30 kN 1 kg 10 N 10 kg 2 kg 5 kg PCE DDM 5 5000 kg 50 kN 2 kg 20 N 40 kg 3...

Page 6: ...mm B mm C mm D mm mm H mm Material PCE DDM 3 260 123 37 195 51 365 aluminium PCE DDM 5 285 123 57 210 58 405 aluminium PCE DDM 10 320 120 57 230 92 535 steel PCE DDM 20 375 128 74 260 127 660 steel PC...

Page 7: ...ta memory f STB stable condition g kN kilonewton N newton in taring mode the N flashes h t tonne i lb lbs k kg kilogrammes 5 1 2 Keypad a ON OFF 1 Press the ON OFF key to switch the device on 2 When t...

Page 8: ...to measurement mode Press MODE 4 times The display will show Unit Then press ENTER to get to the mode in which you can select the unit Press Zero to select the unit and Mode to return to measurement...

Page 9: ...switches off automatically after 60 minutes 5 3 2 Zero OPERATION DISPLAY When the device is switched on it normally displays In case it shows small numbers press this key 1 Press ZERO 2 Press ZERO 5 3...

Page 10: ...calibration You can use the hand held device to select parameter settings or calibration 5 4 1 Key functions Key Function Tare to change numbers ZERO to increase values MODE to save and exit ENTER to...

Page 11: ...rated it will display the calibrated value Press Zero to select Select the unit Un 0 kg Un 1 lb Un 2 N Un 3 KN Un 4 t Normally select Un 0 Press Enter to confirm If the device is new it will display O...

Page 12: ...after attaching the tare weight You can also attach the tare weight before switching the device on Press MODE 8 times The display will show CLIBr Press ENTER and hold it for 3 seconds to confirm the c...

Page 13: ...attached before STB light flashes 5 6 Troubleshooting problem possible reason solution no display batteries defective replace key defective call support ON OFF key has not been pressed properly press...

Page 14: ...a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 7 Contact If you have any questions about our range of products or measurin...

Reviews: