![PCE Instruments PCE-DDM 10 Manual Download Page 11](http://html1.mh-extra.com/html/pce-instruments/pce-ddm-10/pce-ddm-10_manual_1516102011.webp)
Manual
11
it will display the calibrated
decimal place.
Press “Zero” to select.
Select the decimal place. 0/1/2/3
is displayed one after the other.
0 = xxxxx, 1 = xxxx,x, 2 = xxx,xx,
3 = xx,xxx
The capacity is 2000 kg, so
choose Pt = 0
Press “Enter” to confirm.
Device shows the zero point.
Press “Enter” to confirm.
If the device is new, it will display
Cd = 11. If the device is
calibrated, it will display the
calibrated value.
Press “Zero” to confirm.
C: search zero point. Range: 0 …
5
0 = 0 d, 1 = 0.5 d, 2 = 1 d, 3 = 1.5
d, 4 = 2 d, 5 = 2.5 d
d: display speed 3 options:
0 = slow (1.24 Hz), 1 = medium
(2.5 Hz), 2 = fast (5 Hz)
Press “Enter” to confirm.
If the device is new, it will display
LL = 2. If the device is calibrated,
it will display the calibrated value.
Press “Zero” to select.
LL: filter parameter, 6 options (0
… 5) from low to high.
Select LL = 0 before calibration
and LL = 2 after calibration.
Press “Enter” to confirm.
If the device is new, it will display
Un = 0. If the device is calibrated,
it will display the calibrated value.
Press “Zero” to select.
Select the unit. Un = 0: kg, Un =
1: lb, Un = 2: N, Un = 3: KN, Un =
4: t
Normally select Un = 0.
Press “Enter” to confirm.
If the device is new, it will display
OFF 1. If the device is calibrated,
it will display the calibrated value.
Press “Zero” to select.
Choose the switch-off mode. You
have 3 options: 0. manual switch-
off, 1. remote switch-off, 2.
automatic switch-off
Press “Enter” to confirm.
set point 1
Press “Enter” to confirm.
If the device is new, it will display
1 OFF. If the device is calibrated,
it will display the calibrated value.
Press “Zero” to select.
Select set point 1. “1 OFF” means
that set point 1 does not inherit a
function. “1 LO” means that set
point 1 is 1 lower. “1 HI” means
that set point 2 is 1 higher.
Press “Enter” to confirm.
If the device is new, it will display
02000. If the device is calibrated,
it will display the calibrated value.
Press “ZERO” to determine the
value and “TARE” to move the
decimal place.
Press “Enter” to confirm.
set point 2
Press “Enter” to confirm.
If the device is new, it will display
2 OFF. If the device is calibrated,
it will display the calibrated value.