Press “Zero” to select.
Select set point 2. “2 OFF” means
that set point 2 does not inherit a
function. “1 LO” means that set
point 2 is 1 lower. “1 HI” means
that set point 2 is 1 higher.
Press “TARE” to confirm.
If the device is new, it will display
01200. If the device is calibrated,
it will display the calibrated value.
Press “ZERO” to determine the
value and “TARE” to move the
decimal place.
Press “Enter” to confirm.
Gravity is taken into account.
Press “Enter” to confirm.
If the device is new, it will display
97940. If the device is calibrated,
it will display the calibrated value.
Press “Mode”.
Confirm the above parameter
settings.
5.4.3
Calibration
OPERATION
DISPLAY
Description
Select 2000 kg as capacity and
make sure that the display
shows
after attaching
the tare weight. You can also
attach the tare weight before
switching the device on.
Press “MODE” 8 times. The
display will show “CLIBr”. Press
“ENTER” and hold it for 3
seconds to confirm the
calibration.
no load
The display shows 0. Wait for the
STB signal.
Attach a standard weight of 2000
kg.
Wait until the STB signal is
displayed.
Press the “Enter” key.
If the first number in the
parameter settings flashes,
select the capacity and wait for
the STB signal on the screen.
Press “Mode” to save.
Confirm.
Press “Tara” to move the
decimal place.
Calibration finished.
5.5
Display explanations
DISPLAY
explanation
note
no wireless signal
open parameter settings
open tension settings
calibration point
select max. limit
tare weight cannot be the max.
limit
select min. limit
tare weight cannot be the min.
limit
wait
weight < minimum load
cannot be weighed
saving time exceeded
cannot save after 30 times
memory full
cannot save more than 99999