PCE Instruments PCE-DDM 10 Manual Download Page 12

Manual 

12 

Press “Zero” to select. 

 

Select set point 2. “2 OFF” means 

that set point 2 does not inherit a 

function. “1 LO” means that set 

point 2 is 1 lower. “1 HI” means 

that set point 2 is 1 higher. 

Press “TARE” to confirm. 

 

If the device is new, it will display 
01200. If the device is calibrated, 

it will display the calibrated value. 

Press “ZERO” to determine the 

value and “TARE” to move the 

decimal place. 

 

 

Press “Enter” to confirm. 

 

Gravity is taken into account. 

Press “Enter” to confirm. 

 

If the device is new, it will display 
97940. If the device is calibrated, 

it will display the calibrated value. 

Press “Mode”. 

 

Confirm the above parameter 

settings. 

5.4.3 

Calibration 

OPERATION 

DISPLAY 

Description

 

 

 

Select 2000 kg as capacity and 

make sure that the display 

shows 

after attaching 

the tare weight. You can also 

attach the tare weight before 

switching the device on. 

Press “MODE” 8 times. The 

display will show “CLIBr”. Press 

“ENTER” and hold it for 3 

seconds to confirm the 

calibration. 

 

no load 

The display shows 0. Wait for the 

STB signal. 

Attach a standard weight of 2000 

kg. 

 

Wait until the STB signal is 

displayed. 

Press the “Enter” key. 

 

If the first number in the 

parameter settings flashes, 

select the capacity and wait for 

the STB signal on the screen. 

Press “Mode” to save. 

 

Confirm. 

 

 

 

 

 

 

 

 

 

Press “Tara” to move the 

decimal place. 

 

Calibration finished. 

5.5 

Display explanations 

DISPLAY 

explanation 

note 

 

no wireless signal 

 

 

open parameter settings 

 

 

open tension settings 

 

 

calibration point 

 

 

select max. limit 

tare weight cannot be the max. 

limit 

 

select min. limit 

tare weight cannot be the min. 

limit 

 

wait 

 

 

weight < minimum load 

cannot be weighed 

 

saving time exceeded 

cannot save after 30 times 

 

memory full 

cannot save more than 99999 

Summary of Contents for PCE-DDM 10

Page 1: ...nts UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From Outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instr...

Page 2: ...ide 9 5 3 1 ON OFF 9 5 3 2 Zero 9 5 3 3 TARE 9 5 3 4 PEAK HOLD 9 5 3 5 Data memory 9 5 3 6 Data search 9 5 3 7 Delete memory 10 5 3 8 Battery voltage 10 5 3 9 To change the unit 10 5 3 10 To return to...

Page 3: ...of the measuring equipment must be borne in mind for example those of the crane producer of the load suspension device producer and of the shackle producer In this relation the measurement equipment m...

Page 4: ...ar of the attachment parts Control all dimensions Consider all possible deformations and damage Make sure that all parts work properly Every 12 months or 50000 measurements whatever happens first Insp...

Page 5: ...ies operation time approx 50 h battery life 40 hours model range resolution minimum load weight with shackle PCE DDM 3 3000 kg 30 kN 1 kg 10 N 10 kg 2 kg 5 kg PCE DDM 5 5000 kg 50 kN 2 kg 20 N 40 kg 3...

Page 6: ...mm B mm C mm D mm mm H mm Material PCE DDM 3 260 123 37 195 51 365 aluminium PCE DDM 5 285 123 57 210 58 405 aluminium PCE DDM 10 320 120 57 230 92 535 steel PCE DDM 20 375 128 74 260 127 660 steel PC...

Page 7: ...ta memory f STB stable condition g kN kilonewton N newton in taring mode the N flashes h t tonne i lb lbs k kg kilogrammes 5 1 2 Keypad a ON OFF 1 Press the ON OFF key to switch the device on 2 When t...

Page 8: ...to measurement mode Press MODE 4 times The display will show Unit Then press ENTER to get to the mode in which you can select the unit Press Zero to select the unit and Mode to return to measurement...

Page 9: ...switches off automatically after 60 minutes 5 3 2 Zero OPERATION DISPLAY When the device is switched on it normally displays In case it shows small numbers press this key 1 Press ZERO 2 Press ZERO 5 3...

Page 10: ...calibration You can use the hand held device to select parameter settings or calibration 5 4 1 Key functions Key Function Tare to change numbers ZERO to increase values MODE to save and exit ENTER to...

Page 11: ...rated it will display the calibrated value Press Zero to select Select the unit Un 0 kg Un 1 lb Un 2 N Un 3 KN Un 4 t Normally select Un 0 Press Enter to confirm If the device is new it will display O...

Page 12: ...after attaching the tare weight You can also attach the tare weight before switching the device on Press MODE 8 times The display will show CLIBr Press ENTER and hold it for 3 seconds to confirm the c...

Page 13: ...attached before STB light flashes 5 6 Troubleshooting problem possible reason solution no display batteries defective replace key defective call support ON OFF key has not been pressed properly press...

Page 14: ...a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 7 Contact If you have any questions about our range of products or measurin...

Reviews: