Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
14
1-800-343-4048
6
Install the new high pressure (shorter one) tip seal starting from the center of the scroll, outward
inside the seal channel. (Figures 25 - 27)
Note:
The side and bottom lip notches of the tip seal face INWARD and DOWNWARD
in the channel.
7
Install the new low pressure (longer one) tip seal in the same way. Note: Make sure there is
no gap between the high and low pressure seals. (Figures 28, 29)
Note:
Make sure there is no gap between the high and low pressure seals.
Caution! Do not attempt to remove the orbit scroll from the housing.
Figure 25
Figure 26
Figure 27
Figure 28
Figure 29
8
After installing half of the low pressure seal,
carefully
remove the seal from the channel to
make sure the seal is properly locking onto the channel indentations located just past the high
pressure seal.
Note:
The indentations are machined into the seal channel to prevent the low pressure seal
from moving.
9
Blow off any dust caused by removing the seal from the channel.
10
Install the low pressure seal completely and cut excessive material. (Figures 30-32)
Note:
Make sure the side and bottom lip notches are facing INWARD and DOWN into the
seal channel.
Note:
The lip notches must not be distorted in the seal or torn off.
11
Install backup tube in the dust seal channel.
Figure 30
Figure 31
Figure 32
HP Tip Seal for OS
HP Tip Seal for FS
LP Tip Seal
for OS
LP Tip Seal
for FS
Dust Seal
Backup Tube