Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
18
1-800-343-4048
System Specifications
Cautions
1
The generator should be installed in an area with adequate ventilation to reduce the
flammability of the oxygen-rich permeate stream. The system should not be located in an
area where the permeate stream poses the risk of explosion or combustion
2
Nitrogen is nontoxic and largely inert. It can act as a simple asphyxiant by displacing
oxygen in air. Inhalation of nitrogen in excessive concentrations can result in
unconsciousness without any warning symptoms such as dizziness and fatigue
.
3
The maximum operating pressure of the system is 137 psig (9.4 barg). Operating the
generator at pressures above 137 psig (9.4 barg) will result in damage to the pump and
membrane.
4
The drain lines should be piped away to the provided collection container (open to
atmospheric pressure) to avoid any possible re-entrainment of liquid in the air which feeds
the generator.
5
Changes in inlet pressure or outlet flow demand will alter the gas purity.
6
Install the generator in a room where the ambient temperature and humidity are below
85°F at 43% RH.
System Specifications
NitroFlow TG2
CSA Certification Standard
CAN/CSA C22.2 No. 61010-1
IEC Standard
IEC 61010-1
UL Standard
UL 61010
IEC 61010 Installation
Category II
IEC 61010 Pollution
Degree 2
Impact Code
IK08
EMC Compliance
EN61326-1/CISPR 11
Curtain Gas (N2) Purity
99.999% Organics, >96->93% N2 depending on flow
Curtain Gas Max. Flow/Pressure
30 SLPM/80 PSIG (5.5 barg)
Gas 1 / Gas 2 Purity
2
<0.1 PPM Total Hydrocarbons / -40°F (-40°C) ATM Dew Point
Gas 1 / Gas 2 Max. Flow Pressure
34 SLPM/110 PSIG (8.2 barg)
Source Gas (Dry Air) Purity
<0.1 PPM Total Hydrocarbons / -40°F (-40°C) ATM Dew Point
Source Gas Max. Flow Pressure
34 SLPM/110 PSIG (8.2 barg)
Suspended Liquids
None
Particles > 0.01 Micron
None
Phthalates None
Sound Level
<49 dB(A) at 1 meter
Min./Max. Compressor Operating Pressure
130/137 PSIG (8.9 to 9.4 barg)
Min./Max. Ambient Temp.
60°F/85°F (16°C-29°C)
Max. Ambient Relative Humidity
43% at 86°F – see chart on page 17
Altitude
2000M (for higher altitude, see chart on page 18)
Heat Dissipation
10246 BTU/Hour
Electrical Requirements¹ - 60 Hz
208-254VAC, 60 Hz, 1 Phase, 16A
Electrical Requirements¹ - 50 Hz
230VAC, 50 Hz, 1 Phase, 13A
Power Consumption - 60 Hz
16A@230VAC, 3680 Watts
Power Consumption - 50 Hz
13A@230VAC, 2990 Watts
Start-Up Time
10 Minutes
Compressor Outlet - Tank Drain
1/4” Tube
Ports - Filter Drains (2)
1/4” Tube
Ports - Water Separator
1/4” Tube
Ports - Nitrogen Outlet
1/4” NPT
Ports - Dry Air Outlet
1/4” NPT
Ports - Hydrocarbon Free Air
1/4” NPT
Fuses - Primary FU1, FU2
10.3 x 38.1mm, Type T, 600V, 1.25A
Fuses - Secondary FU3
6.3 x 31.7mm, Type T, 250V, 1.25A
Dimensions
33.5”L x 20.75”W x 45.25”H
Product Weight
427 lbs (194 kgs)
Shipping Weight
637 lbs (289 kgs)
¹ Main supply voltage fluctuations not to exceed ± 10% of nominal voltage.
2
Hydrocarbons other than methane.