Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
19
1-800-343-4048
Accessory Parts
Part Number
Voltage Step-Down Transformer
A03-0412
Voltage Step-Up Transformer
A03-0411
Exhaust Ventilation Kit
B04-0575
Bottle Drain Kit
B04-0540
In-line Duct Fan (120V)
A03-0378
Exhaust Duct Kit
B04-0693
Temperature (°F)
Relative Humidity (%)
86
43
80
51
75
60
70
71
65
84
60
100
Maximum Relative Humidity at Different Temperatures
Compressor Output Pressure at Different Altitudes
Disconnect the electrical power and depressurize the generator before
attempting any troubleshooting activities according to the shut down steps in
the operation section. Only trained personnel using reasonable care should
perform any troubleshooting activities.
Determine whether the trouble is with the compressor or the gas generator.
Disconnect the plastic tubing from the compressor fitting (see Figure 43). Once
removed, place supplied plug into fitting. Check that the compressed air source
builds full pressure and cycles off within two minutes.
If the compressor air supply is confirmed, remove lower side panel and relieve tank
pressure using auto tank drain test button. Proceed to the NitroFlow TG2 generator
Troubleshooting section (page 20), and begin troubleshooting the NitroFlow TG2
generator. Otherwise, refer to the Compressor Troubleshooting section (pages 19,
20), and begin troubleshooting the compressor.
Troubleshooting
Altitude
Max Compressor Pressure
Max Outlet Pressure
at Full Flow
0 – 3,280 ft (0 – 1000 m)
125 psig
115 psig
3,281 – 8,200 ft (1,001 – 2,500 m)
125 psig
115 psig
8,201 – 13,100 ft (2,501 – 4,000 m)
116 psig
106 psig
13,101 ft (4,001 m)
N/A
N/A
Figure 48