background image

C.  Getting Started 

5                                     Questions about eRapid?   Call 1-866-335-6943 

Getting Started 

 

 
 
 
 
 

Check your eRapid System package to make sure you have the items shown below.  Note that the package contains 

two eRapid Handsets and two eRapid Aerosol Heads.  Also note that the 

eRapid Handset is not sterile and needs to 

be cleaned and/or disinfected before you use it for the first time

.  Remove one new Handset and one new Aerosol 

Head from their packaging and proceed to step E2c to begin the cleaning procedure.  You can keep the second 

Handset as a backup or clean it at the same time as you clean the first Handset.

 

eBase Controller 

Connection Cord 

Carrying Case 

AC Power Supply 

4 “AA” Batteries 

Easycare Cleaning Aid 

Blue Valve 

Aerosol Head 

Medication Cap 

Medication Reservoir 

Aerosol Chamber 

Mouthpiece with 

Blue Flap 

Summary of Contents for eRapid

Page 1: ...General Purpose Electronic Nebulizer...

Page 2: ...ng Handset components in warm soapy water for 5 minutes 3 Rinse all parts 4 Disinfect Handset components by boiling in distilled water for 5 minutes 5 Air dry Handset components until dry 3 Assemble a...

Page 3: ...lean by soaking Handset components in warm soapy water for 5 minutes 3 Rinse all components to remove residual soap and inspect for dirt Soak for an additional 5 minutes if necessary 4 Disinfect by bo...

Page 4: ...d Nebulizer System read and understand all Instructions For Use Save these Instructions For Use for future reference If for any reason you Do Not understand any part of these Instructions For Use plea...

Page 5: ...ow when the Controller is damaged or not properly assembled 12 To reduce the risk of infection illness and or injury from contamination and to maintain the longevity of your Handset clean and dry all...

Page 6: ...be cleaned and or disinfected before you use it for the first time Remove one new Handset and one new Aerosol Head from their packaging and proceed to step E2c to begin the cleaning procedure You can...

Page 7: ...output Operating life is about 2 hours when using high quality disposable batteries To reduce the risk of battery leakage it is recommended to remove the batteries if you plan to store the Controller...

Page 8: ...ishes Use only distilled water to disinfect by boiling Follow manufacturer directions when using a chemical disinfectant such as Control III Handset Maintenance Summary Details Can Be Found on the Fol...

Page 9: ...rformance It is important to rinse and clean your Handset right away after every use Step E1 Disconnect the Handset from the Connection Cord Figure E1 Step E2 Disassemble the eRapid Handset a Remove t...

Page 10: ...Figure E4 Never use a brush or any other object to clean the Aerosol Head Thorough rinsing of both sides of the Aerosol Head helps with proper operation and prevents clogging Step E5 Soak all parts of...

Page 11: ...lled water for a full 5 minutes Figure E9b Step E10 Dry on a lint free cloth or paper towel Figure E10 After the parts are completely dry place them in a bag or cover them with a lint free cloth or pa...

Page 12: ...ce Step E13 Connect the Easycare to the Controller using the Connection Cord by attaching the horseshoe shaped end of the Connection Cord upside down to the hinge of the Easycare part and swivel it do...

Page 13: ...y pressing and holding the On Off button for 10 seconds Step E17 Reverse the assembly steps and discard any solution from the Easycare Step E18 Clean the Easycare by soaking in warm soapy water Step E...

Page 14: ...re clean and not visibly damaged Do Not use dirty or damaged parts Step F3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab...

Page 15: ...it It is important that you Do Not touch the metal part of the Aerosol Head Figure F5c Step F6 Close the Handset by pushing the Aerosol Chamber together with the Medication Reservoir until you hear o...

Page 16: ...id Do not exceed the maximum fill volume of 6 mL Step G4 Close the Medication Cap by lining up the Tabs on the Medication Cap with the Tab Slots on the Medication Reservoir and turning the Medication...

Page 17: ...Flap with your mouth Step G8 Turn on the Controller Press and hold the On Off button for 1 2 seconds Figure G8a The following will happen You will hear 1 beep and the status light will turn green The...

Page 18: ...opening Figure G11 To mimic other general purpose nebulizers the eRapid is designed to keep 1 mL 1 5 teaspoon of medicine in the Medication Reservoir Do not tilt or try to use the residual medicine If...

Page 19: ...he Batteries when storing the Controller for more than 30 days The Handset should last for 2 6 months but it depends on the number of different medicines used with the Handset the number of times the...

Page 20: ...s blinking No medicine Add medicine if you have not done so Not level or medicine is not touching the Aerosol Head Gently tap the Handset hold the Handset level and press the On Off button 7 Blinking...

Page 21: ...ndition persists call the number below 16 Longer than normal nebulization time Clean the Aerosol Head and then use Easycare If condition persists call the number below 17 There is more than 1 mL resid...

Page 22: ...01 1 Type of electric shock protection AC Power Supply Protection Class II Degree of protection from electric shock of part used nebulizer Type BF Degree of protection against water ingress per IEC 60...

Page 23: ...ent should fail during the warranty period PARI at its option will repair or replace the equipment or parts at issue provided the claim is bona fide It shall be the responsibility of the purchaser to...

Page 24: ...n Virginia 23112 USA 2019 PARI Pharma GmbH All rights reserved The PARI Pharma logo is a registered trademark of PARI GmbH Lamira eBase Altera Zirela eFlow and the eFlow TECHNOLOGY logo are registered...

Reviews: