background image

E.  How To Clean and Maintain Your eRapid 

  10                                      Questions about eRapid?   Call 1-866-335-6943 

 

Disinfecting Your Handset 

 

To reduce the risk of serious or fatal illness or injury caused by 

contamination, it is important to disinfect your Handset daily or as 

recommended by your physician.  Disinfection is more effective if 

you clean your Handset before you disinfect. 

 

Step E9:

 

Disinfect the Handset 

with the commercially 

available medical device disinfectant Control III® or 

by boiling in distilled water.

 

 

 

To disinfect with disinfectant

, mix ½ 

tablespoon of Control III solution to 1 quart of 

water in a clean bowl.  Immerse the parts in that 

bowl and soak for 

10 minutes

 (

Figure E9a

).  

Remove parts from solution and rinse thoroughly 

with sterile water. 

 

 

To disinfect with boiling water

, immerse and 

boil the parts, including the Aerosol Head, in a 

clean pot of 

distilled

 water for a full 5 minutes 

(

Figure E9b

). 

 
 

Step E10:

 

Dry on a lint-free cloth or paper towel (Figure 

E10)

.  After the parts are completely dry, place 

them in a bag or cover them with a lint-free cloth or 

paper towel for storage. Re-assemble the Handset 

just prior to the next treatment.   

 

Unclogging the Aerosol Head with Easycare 

(Optional) 

 

After multiple uses, your nebulization time may slowly increase.  If 

soaking in soapy water does not improve performance, it may be 

necessary to perform special cleaning of the Aerosol Head.  The 

Easycare

 is used to clean and unclog the small holes in the 

Aerosol Head by reversing the flow of liquid.  The cleaning process 

takes about 5-10 minutes and uses the 

Easycare

 powered by the 

Controller. 

 It is important NOT TO USE 

the 

Easycare

 more than 

30 times on a single Aerosol Head. 

Clean the Handset as described in Steps E1-E7.  

Step E11:

 

Open the Easycare.  

 

Press lightly on the side gripping surfaces to release 

the side engaging hooks

 (Figure E11)

.    

 

Gently pull the light gray funnel part outward and 

down. 

 

E11

 

E9a

 

E9b

 

E10

 

Summary of Contents for eRapid

Page 1: ...General Purpose Electronic Nebulizer...

Page 2: ...ng Handset components in warm soapy water for 5 minutes 3 Rinse all parts 4 Disinfect Handset components by boiling in distilled water for 5 minutes 5 Air dry Handset components until dry 3 Assemble a...

Page 3: ...lean by soaking Handset components in warm soapy water for 5 minutes 3 Rinse all components to remove residual soap and inspect for dirt Soak for an additional 5 minutes if necessary 4 Disinfect by bo...

Page 4: ...d Nebulizer System read and understand all Instructions For Use Save these Instructions For Use for future reference If for any reason you Do Not understand any part of these Instructions For Use plea...

Page 5: ...ow when the Controller is damaged or not properly assembled 12 To reduce the risk of infection illness and or injury from contamination and to maintain the longevity of your Handset clean and dry all...

Page 6: ...be cleaned and or disinfected before you use it for the first time Remove one new Handset and one new Aerosol Head from their packaging and proceed to step E2c to begin the cleaning procedure You can...

Page 7: ...output Operating life is about 2 hours when using high quality disposable batteries To reduce the risk of battery leakage it is recommended to remove the batteries if you plan to store the Controller...

Page 8: ...ishes Use only distilled water to disinfect by boiling Follow manufacturer directions when using a chemical disinfectant such as Control III Handset Maintenance Summary Details Can Be Found on the Fol...

Page 9: ...rformance It is important to rinse and clean your Handset right away after every use Step E1 Disconnect the Handset from the Connection Cord Figure E1 Step E2 Disassemble the eRapid Handset a Remove t...

Page 10: ...Figure E4 Never use a brush or any other object to clean the Aerosol Head Thorough rinsing of both sides of the Aerosol Head helps with proper operation and prevents clogging Step E5 Soak all parts of...

Page 11: ...lled water for a full 5 minutes Figure E9b Step E10 Dry on a lint free cloth or paper towel Figure E10 After the parts are completely dry place them in a bag or cover them with a lint free cloth or pa...

Page 12: ...ce Step E13 Connect the Easycare to the Controller using the Connection Cord by attaching the horseshoe shaped end of the Connection Cord upside down to the hinge of the Easycare part and swivel it do...

Page 13: ...y pressing and holding the On Off button for 10 seconds Step E17 Reverse the assembly steps and discard any solution from the Easycare Step E18 Clean the Easycare by soaking in warm soapy water Step E...

Page 14: ...re clean and not visibly damaged Do Not use dirty or damaged parts Step F3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab...

Page 15: ...it It is important that you Do Not touch the metal part of the Aerosol Head Figure F5c Step F6 Close the Handset by pushing the Aerosol Chamber together with the Medication Reservoir until you hear o...

Page 16: ...id Do not exceed the maximum fill volume of 6 mL Step G4 Close the Medication Cap by lining up the Tabs on the Medication Cap with the Tab Slots on the Medication Reservoir and turning the Medication...

Page 17: ...Flap with your mouth Step G8 Turn on the Controller Press and hold the On Off button for 1 2 seconds Figure G8a The following will happen You will hear 1 beep and the status light will turn green The...

Page 18: ...opening Figure G11 To mimic other general purpose nebulizers the eRapid is designed to keep 1 mL 1 5 teaspoon of medicine in the Medication Reservoir Do not tilt or try to use the residual medicine If...

Page 19: ...he Batteries when storing the Controller for more than 30 days The Handset should last for 2 6 months but it depends on the number of different medicines used with the Handset the number of times the...

Page 20: ...s blinking No medicine Add medicine if you have not done so Not level or medicine is not touching the Aerosol Head Gently tap the Handset hold the Handset level and press the On Off button 7 Blinking...

Page 21: ...ndition persists call the number below 16 Longer than normal nebulization time Clean the Aerosol Head and then use Easycare If condition persists call the number below 17 There is more than 1 mL resid...

Page 22: ...01 1 Type of electric shock protection AC Power Supply Protection Class II Degree of protection from electric shock of part used nebulizer Type BF Degree of protection against water ingress per IEC 60...

Page 23: ...ent should fail during the warranty period PARI at its option will repair or replace the equipment or parts at issue provided the claim is bona fide It shall be the responsibility of the purchaser to...

Page 24: ...n Virginia 23112 USA 2019 PARI Pharma GmbH All rights reserved The PARI Pharma logo is a registered trademark of PARI GmbH Lamira eBase Altera Zirela eFlow and the eFlow TECHNOLOGY logo are registered...

Reviews: