B.  Safety Precautions 

4                                     Questions about eRapid?   Call 1-866-335-6943 

Read the following Dangers, Warnings, and 

Cautions before using this device. 

 

 

To reduce the risk of serious or fatal injury from electrocution: 

1.

 

Do Not

 place or store your eRapid where it can be in water or fall into a 

bathtub, sink, or other liquid.  

Do Not

 place or drop into water or other 

liquid. 

Do Not

 use while bathing. 

2.

 

Do Not

 reach for the eRapid if it has fallen into water or other liquid. 

Unplug immediately.  Retrieve your eRapid 

ONLY

 after it has been 

unplugged.  

 

 

1.

 

The eRapid Nebulizer Handset (Handset) is for use only by 1 patient.  

Do Not

 

Share

 your Handset with other people. 

2.

 

The eRapid Handset has a residual volume similar to general purpose 

compressor-based jet nebulizers and will leave about 1/5 teaspoon (1 

mL) of medicine in the reservoir

.  Do not use this residual medicine 

or your dosage will not be accurate.

 

3.

 

Read, understand and follow all warnings and instructions in the 

Instructions For Use prior to using this device. 

4.

 

To reduce the risk of damage and malfunction of the unit that could 

result in serious or fatal injury from electrocution, fire, and burns: 

 

a.

  

Do Not

 overload wall outlets or use extension cords. 

 

b

.  Keep all electrical cords away from heated surfaces. 

 

c.

  

Do Not 

spray liquids onto the housing of the eBase® Controller 

(Controller).  See Section C: Getting Started. Liquid may cause 

damage to the electrical parts and could lead to a malfunction. In 

the event that liquids enter the Controller, call the number below. 

 

d.

  

Do Not

 operate where oxygen is being used in a closed area such 

as an oxygen tent. 

 

e.

  Always unplug this product immediately after using and before 

cleaning.

 

5.

 

Use

 

Only

 AC power supplies and accessories that are authorized by 

PARI.  Use of unapproved power supplies or accessories can lead to 

improper treatment, injury, or cause (receive) interference leading to 

damage to the Controller. 

6.

 

Never

 operate this product if:  

 

a.

  it has damaged cords or plugs, 

 

b.

  it is not working properly, 

 

c.

  it has been dropped or damaged,  

 

d.

  it has been exposed to any liquids inside the Controller, or 

e.

  the AC power supply is visibly damaged. 

7.

 

This product contains small parts that may present a choking hazard to 

small children and the Connection Cord (Connection Cord) may present 

a strangulation hazard. Close supervision is necessary when this 

product is used near children or the physically or mentally impaired. 

8.

 

This Handset has unique performance characteristics and should 

Never

 

be substituted for other eFlow Technology Handsets (e.g. Altera

®

Lamira

®

, Zirela®, eRapid

®

) or Investigational eFlow Handsets. 

9.

 

Never

 use the eRapid Aerosol Head (Aerosol Head) with any other 

eFlow Technology Handset or eFlow Technology System (e.g. Altera, 

Lamira, Zirela, Investigational eFlow).  

10.

 

Do Not

 use your eRapid while driving or in any situation which takes 

away your full attention. 

11.

 

Never

 operate the Controller if it is not properly and completely 

assembled. 

Do Not

 use if the Controller is damaged.  See Section I: 

Troubleshooting, for the alerts that show when the Controller is 

damaged or not properly assembled. 

12.

 

To reduce the risk of infection, illness, and/or injury from contamination, 

and to maintain the longevity of your Handset, clean and dry all parts of 

the Handset before you first use it and after each treatment.  Follow the 

instructions in Section E for cleaning your Handset. 

13.

 

Cleaning the Handset only removes the medicine and saliva.  To reduce 

the risk of serious or fatal illness caused by contamination of the 

Handset, 

disinfect the Handset daily or as recommended by your 

physician

.   See steps E9 and E10 for disinfection instructions. 

14.

 

Only

 use cleaning and disinfecting procedures described in this 

document.  PARI has tested these procedures specifically for eRapid.  

Use of other methods may not clean or disinfect properly or damage 

your Handset. 

 

1.

 

Before use, check your Handset for proper assembly.  All parts must be 

connected and firmly in place. Use of an improperly assembled Handset 

could reduce or eliminate the effectiveness of the treatment. 

2.

 

Keep the Handset level when in use.  If the Handset is tilted, the dosage 

may not be accurate. Excessive tilting can cause the Controller to shut 

off prematurely because medicine may not be covering the Aerosol 

Head membrane. 

3.

 

To reduce the risk of damage and malfunction of the unit,

 Do Not

 drop 

or insert any object into any opening of the eRapid. 

4.

 

Do Not

 modify this equipment without authorization from the 

manufacturer.

 

 

5.

 

Do Not

 disassemble the Controller at any time. There are no user 

serviceable parts inside the Controller. Call the number below for all 

service needs. 

6.

 

Do Not

 allow pets (for example; dogs, rodents, rabbits) near the cables. 

7.

 

Cleaning the Handset properly will help prevent the Aerosol Head from 

clogging.  If the Aerosol Head becomes clogged, the aerosol mist will be 

reduced or stopped, which may make the treatment ineffective.  If 

clogging occurs, use 

Easycare

 and water to unclog the Aerosol Head 

as described in Steps E11 through E19.  

8.

 

If you suspect that your eRapid has been damaged or is not working 

correctly, call the number below. 

9.

 

Do Not

 use the device in areas exposed to elevated electromagnetic or 

electrical radiation such as a MRI scanner or high frequency surgical 

equipment. 

10.

 

Do Not

 place near other medical devices during operation unless both 

devices are monitored constantly to insure both are operating properly. 

11.

 

Do Not

 use within 30 cm of portable wireless communication devices 

such as cell phones or antenna cables or external antennas. 

12.

 

Do Not

 use near airplane or train control systems. Do not use aboard 

aircraft. 

13.

 

Do Not

 use the nebulizer near anti-theft systems and Radio Frequency 

Identification (RFID) readers, which are used in a wide variety of 

settings, including stores, libraries, and hospitals. 

Do not 

power on the 

nebulizer when passing through security screening or theft protection 

(RFID) systems at entrances or exits of stores, libraries, or hospitals.  

Note that some entrance and exit security systems are not visible. 

Technical electromagnetic compatibility data is available in table form upon 

request from PARI Respiratory Equipment or on the Internet at 

https://www.pari.com/fileadmin/Electromagnetic-compatibility-4.pdf 

     

WARNING 

     

DANGER 

     

CAUTION 

Summary of Contents for eRapid

Page 1: ...General Purpose Electronic Nebulizer...

Page 2: ...ng Handset components in warm soapy water for 5 minutes 3 Rinse all parts 4 Disinfect Handset components by boiling in distilled water for 5 minutes 5 Air dry Handset components until dry 3 Assemble a...

Page 3: ...lean by soaking Handset components in warm soapy water for 5 minutes 3 Rinse all components to remove residual soap and inspect for dirt Soak for an additional 5 minutes if necessary 4 Disinfect by bo...

Page 4: ...d Nebulizer System read and understand all Instructions For Use Save these Instructions For Use for future reference If for any reason you Do Not understand any part of these Instructions For Use plea...

Page 5: ...ow when the Controller is damaged or not properly assembled 12 To reduce the risk of infection illness and or injury from contamination and to maintain the longevity of your Handset clean and dry all...

Page 6: ...be cleaned and or disinfected before you use it for the first time Remove one new Handset and one new Aerosol Head from their packaging and proceed to step E2c to begin the cleaning procedure You can...

Page 7: ...output Operating life is about 2 hours when using high quality disposable batteries To reduce the risk of battery leakage it is recommended to remove the batteries if you plan to store the Controller...

Page 8: ...ishes Use only distilled water to disinfect by boiling Follow manufacturer directions when using a chemical disinfectant such as Control III Handset Maintenance Summary Details Can Be Found on the Fol...

Page 9: ...rformance It is important to rinse and clean your Handset right away after every use Step E1 Disconnect the Handset from the Connection Cord Figure E1 Step E2 Disassemble the eRapid Handset a Remove t...

Page 10: ...Figure E4 Never use a brush or any other object to clean the Aerosol Head Thorough rinsing of both sides of the Aerosol Head helps with proper operation and prevents clogging Step E5 Soak all parts of...

Page 11: ...lled water for a full 5 minutes Figure E9b Step E10 Dry on a lint free cloth or paper towel Figure E10 After the parts are completely dry place them in a bag or cover them with a lint free cloth or pa...

Page 12: ...ce Step E13 Connect the Easycare to the Controller using the Connection Cord by attaching the horseshoe shaped end of the Connection Cord upside down to the hinge of the Easycare part and swivel it do...

Page 13: ...y pressing and holding the On Off button for 10 seconds Step E17 Reverse the assembly steps and discard any solution from the Easycare Step E18 Clean the Easycare by soaking in warm soapy water Step E...

Page 14: ...re clean and not visibly damaged Do Not use dirty or damaged parts Step F3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab...

Page 15: ...it It is important that you Do Not touch the metal part of the Aerosol Head Figure F5c Step F6 Close the Handset by pushing the Aerosol Chamber together with the Medication Reservoir until you hear o...

Page 16: ...id Do not exceed the maximum fill volume of 6 mL Step G4 Close the Medication Cap by lining up the Tabs on the Medication Cap with the Tab Slots on the Medication Reservoir and turning the Medication...

Page 17: ...Flap with your mouth Step G8 Turn on the Controller Press and hold the On Off button for 1 2 seconds Figure G8a The following will happen You will hear 1 beep and the status light will turn green The...

Page 18: ...opening Figure G11 To mimic other general purpose nebulizers the eRapid is designed to keep 1 mL 1 5 teaspoon of medicine in the Medication Reservoir Do not tilt or try to use the residual medicine If...

Page 19: ...he Batteries when storing the Controller for more than 30 days The Handset should last for 2 6 months but it depends on the number of different medicines used with the Handset the number of times the...

Page 20: ...s blinking No medicine Add medicine if you have not done so Not level or medicine is not touching the Aerosol Head Gently tap the Handset hold the Handset level and press the On Off button 7 Blinking...

Page 21: ...ndition persists call the number below 16 Longer than normal nebulization time Clean the Aerosol Head and then use Easycare If condition persists call the number below 17 There is more than 1 mL resid...

Page 22: ...01 1 Type of electric shock protection AC Power Supply Protection Class II Degree of protection from electric shock of part used nebulizer Type BF Degree of protection against water ingress per IEC 60...

Page 23: ...ent should fail during the warranty period PARI at its option will repair or replace the equipment or parts at issue provided the claim is bona fide It shall be the responsibility of the purchaser to...

Page 24: ...n Virginia 23112 USA 2019 PARI Pharma GmbH All rights reserved The PARI Pharma logo is a registered trademark of PARI GmbH Lamira eBase Altera Zirela eFlow and the eFlow TECHNOLOGY logo are registered...

Reviews: