311-7001200-035 

Version 4.0 2017/11

1. Mouthpiece and mask are intended for use by one user. DO NOT SHARE.
2. Accessories are consumable parts, e.g. mesh cap. Clean the accessories after each use and 
    replace with new accessories when they become dirty and obstructed.
3. Mask and AC Adapter may be provided separately, please contact your local dealer for 
    purchase, or contact our customer service line for more information.

NOTE

Do not wet the connector and connector socket (refer to Description of Part 3 & Part 4).

Use the equipment only for its intended use as described in the operating instructions. Do not use 
attachments not recommended by the manufacturer. 

Use the AC Adapter (Input: AC, 100-240V 

& Output: DC, 6V) provided by ForaCare, Inc.

Do not use the device under conditions of inter-hospital transport. This device is not intended for 
transport use.
Use the equipment with medications only under the instruction of your physician.
Do not use the equipment if it has any damaged parts, or it has fallen into water.
Prior to using the equipment for the first time, or after storing it for an extended period, be sure to 
clean all necessary parts as described in the cleaning instructions.
Do not use while bathing.
The unit should not be left unattended while plugged in or turned on.
Keep the unit out of reach of small unsupervised children. The small parts detached from the device 
may be a choking hazard if inhaled or swallowed.
To prevent any dangers or hazards, please do not try to modify the device in any way.
Do not place the device in liquid, nor put it where it could fall into liquid. If the device becomes 
wet, unplug the device before touching it.
Do not use the device if it is not working properly, or if it has suffered any damage.
Avoid dropping, smashing, hitting, or tapping the nebulizer unit or the mesh cap.
Please clean the nebulizer accessories after every use, or after storing for a period of time.
Do not expose the nebulizer to direct sunlight, high temperatures or humidity.
Performance information provided by the manufacturer in accordance with the standard EN 
13544-1:2007 may not apply to drugs supplied in suspension or high viscosity form. Please consult 
the drug supplier for such information.

IMPORTANT INFORMATION ABOUT INHALATION THERAPY

SAFETY PRECAUTIONS

The device is ideally suitable for inhalation at home or when traveling. 
Carry out inhalation in a quiet and relaxed state, and inhale slowly and deeply, so that the medication 
can penetrate to the fine, deep bronchial tubes. Exhale normally.

Ensure that all parts are clean and dry.
Before starting treatment, talk to your physician about the duration, dosage and frequency of use 
of the nebulizer.
The indicator light will turn green when the device is working normally.
The device automatically turns off when idle for 15 minutes.

2. Attach the mesh   
    cap and screw it 
    tightly in clockwise
    direction.

1. Press and release the 
    ON/OFF button. The unit
    turns on and delivers 
    prescribed medication
    continuously for up to 15 
    minutes.

4. Attach the mask or the mouthpiece to the adapter.
5. Connect the AC adapter or go to step 6 if the batteries are 
    already installed.
6. Press the ON/OFF button to begin treatment according to 
    the physician’s instructions. See "Delivery Mode" for details 
    on automatic and manual deliveries.
7. To end the treatment, press the ON/OFF button to turn off.

OPERATING INSTRUCTIONS

Important information before use.  These operating instructions provide important information to 
help you use the device properly.

Unit Assembly and Usage

DELIVERY MODE

CLEANING AND MAINTENANCE

This system provides you with two delivery modes: automatic or manual.

Cleaning of Nebulizer Accessories 

Please clean your device accessories after each use.
Prior to cleaning, make sure you have some 

distilled water

 on hand and ensure that the device has 

completed the last medicine delivery cycle and is completely turned off. 

* AUTOMATIC AEROSOL DELIVERY

FEATURES

CONTENTS INCLUDED & DESCRIPTIONS

INTENDED USE

Delivers aerosol automatically or manually
Battery powered or AC adapter connection (optional) 
Small nebulization particle size – approx. 5 microns
Convenient design – compact and lightweight.  Ideal for traveling.
Short treatment time – with high nebulization rate

Dear SootheNeb NBL100 Owner:

Thank you for purchasing the SootheNeb NBL100, before using this product, please read the following 
contents thoroughly and carefully.

If you have other questions regarding this product, please contact the place of purchase or call our 
customer service line at 888.307.8188.

SootheNeb NBL100 is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize 
physician-prescribed solution for inhalation by the patient, 

except for Pentamidine

. The SootheNeb 

NBL100 is intended for use by patients in a home care environment.

Cautions: Federal law restricts this device to only be sold by or on the order of a physician.

Before you start to use your SootheNeb NBL100, 
please check that the following contents are included 
in the package and not broken. The use of other 
contents not listed may constitute a hazard. 

Mesh Nebulizer

Operating Instructions

NBL100

1.

Unit Lid
Mesh Cap
Connector
Connector Socket
Indicator

2.

3.

5.

4.

6.

Mask or Mouthpiece Adapter

7.

8.

ON/OFF Button

10.

Medication Cup

9.

Battery Cover

11.

13.

2 “AA” Alkaline Batteries 

14.

15.

Mouthpiece

AC Adapter

Mask

OPTIONAL PARTS**

INCLUDED PARTS

12.

2.

1.

4.

3.

5.

6.

7.

8.

10.

9.

11.

15.

14.

13.

12.

Nebulizer Unit
AC Adapter Jack

1. Pour the prescribed 
    medication slowly into 
    the medication cup. 
    The maximum capacity 
    is 5 ml.

3. Attach the mask or 
    mouthpiece adapter.

2. Press the ON/OFF  
    button to switch off 
    the device and
    terminate the 
    treatment.

1. Press and hold the 
    ON/OFF button for 2 
    seconds or longer to 
    begin medication 
    delivery.

1. Remove the Mask or Mouthpiece and remove the Mask/Mouthpiece adapter from the Main Unit. 
    Rinse thoroughly and set aside to dry.  Then remove the Mesh Cap counterclockwise from the 
    Medication Cup and keep the medication cup attached.

To deliver medication manually, press the ON/OFF button intermittently to accompany your breathing
rate.

* MANUAL AEROSOL DELIVERY

Repeat step 1 and 2 for additional medication deliveries.

2. Release the ON/OFF 
    button to stop 
    medication delivery.