Logic
External output
Output operation
OUT1
OUT2
OUT3
Logic
Judgme
nt 1
Judgme
nt 2
Alarm
When N.O. is set
Output operation
Indeterminate
Below the
lower limit of
measurement
range
Within GO
range
Above the
upper limit of
measurement
range
Alarm
Output 1
Judgment 1
OFF
ON
OFF
ON
ON
Output 2
Judgment 2
OFF
OFF
ON
ON
ON
Output 3
Alarm
ON
ON
ON
ON
OFF
When N.C. is set
Output operation
Indeterminate
Below the
lower limit of
measurement
range
Within GO
range
Above the
upper limit of
measurement
range
Alarm
Output 1
Judgment 1
ON
OFF
ON
OFF
OFF
Output 2
Judgment 2
ON
ON
OFF
OFF
OFF
Output 3
Alarm
ON
ON
ON
ON
OFF
Logical output has alarm-specific output. (Output 3) Furthermore, logical output makes LOW/
HIGH identifiable in logical operation of judgment 1 and judgment 2.
7.2 Advanced Settings (
)
WUME-HGT-6
7-25
Summary of Contents for HG-T Series
Page 2: ... MEMO 2 WUME HGT 6 ...
Page 28: ... MEMO 2 8 WUME HGT 6 ...
Page 96: ... MEMO 5 50 WUME HGT 6 ...
Page 99: ... Note 2 Displayed only during calculation 6 1 Flow of Measurement Data WUME HGT 6 6 3 ...
Page 128: ... MEMO 6 32 WUME HGT 6 ...
Page 158: ... No timer OFF 7 2 Advanced Settings 7 30 WUME HGT 6 ...
Page 177: ...Example Positioning a glass plate Setting method 7 2 Advanced Settings WUME HGT 6 7 49 ...
Page 218: ...8 2 3 Side View Attachment HG TSV10 Units mm 8 2 Dimension Drawings 8 14 WUME HGT 6 ...
Page 221: ...10 Troubleshooting 10 1 Troubleshooting 10 2 10 2 Error Messages 10 5 WUME HGT 6 10 1 ...
Page 229: ...Menu Structure General Function Display WUME HGT 6 App 3 ...
Page 231: ...Menu Structure Extended Function Display WUME HGT 6 App 5 ...
Page 232: ... MEMO App 6 WUME HGT 6 ...
Page 235: ... MEMO WUME HGT 6 ...
Page 236: ...Panasonic Industrial Devices SUNX Co Ltd 2020 PRINTED IN JAPAN WUME HGT 6 ...