● The following states are judged to be indeterminate states and displayed as "
".
• Full light interception state
• Full light entry state
5.4.4 Inner Diameter/Gap Detection Mode (
)
This mode measures the inner diameter or gap of the measured object.
On the basis of the TOP side, the distance from the edge switching from light interception to
light entry that is nearer to the TOP side to the next edge switching from light entry to light
interception is detected as the measured value.
There is no need to set the measurement direction.
● The following states are judged to be indeterminate states and displayed as "
".
• Full light interception state
• Full light entry state
5.4.5 Central Position Detection Mode (
)
When an object such as a pin is measured, this mode measures the distance from either the
TOP or BOTTOM side edge to the center of the measured object.
5.4 Operation Mode
WUME-HGT-6
5-25
Summary of Contents for HG-T Series
Page 2: ... MEMO 2 WUME HGT 6 ...
Page 28: ... MEMO 2 8 WUME HGT 6 ...
Page 96: ... MEMO 5 50 WUME HGT 6 ...
Page 99: ... Note 2 Displayed only during calculation 6 1 Flow of Measurement Data WUME HGT 6 6 3 ...
Page 128: ... MEMO 6 32 WUME HGT 6 ...
Page 158: ... No timer OFF 7 2 Advanced Settings 7 30 WUME HGT 6 ...
Page 177: ...Example Positioning a glass plate Setting method 7 2 Advanced Settings WUME HGT 6 7 49 ...
Page 218: ...8 2 3 Side View Attachment HG TSV10 Units mm 8 2 Dimension Drawings 8 14 WUME HGT 6 ...
Page 221: ...10 Troubleshooting 10 1 Troubleshooting 10 2 10 2 Error Messages 10 5 WUME HGT 6 10 1 ...
Page 229: ...Menu Structure General Function Display WUME HGT 6 App 3 ...
Page 231: ...Menu Structure Extended Function Display WUME HGT 6 App 5 ...
Page 232: ... MEMO App 6 WUME HGT 6 ...
Page 235: ... MEMO WUME HGT 6 ...
Page 236: ...Panasonic Industrial Devices SUNX Co Ltd 2020 PRINTED IN JAPAN WUME HGT 6 ...