2-47
Sensitivity Adjustment Procedure (In case of type A, D, E, F, G, I and J Models)
1. Connect the each clips of LISTA cable to test point.
2. Set the adjustment mode according to adjustment heads.
(Refer to [LISTA Adjustment Modes] of the “Table of Test Point”.)
3. Playback the LISTA Alignment Tape.
4. Select the item “<6> ATF Error Signal Monitor” on the LISTA main menu, and after appeared message ”1.2%
Speed. . .” on the screen, press the Enter key, then sensitivity value as real time and waveform appear on the
screen as shown below figure.
5. Adjust EVR so that the “Sens. Value” on the upper right corner of the screen is within specification.
6. After finish this adjustment, press the ESC key to return to the main menu.
Adjust within specification