Chapter 4. Error Codes
94
Sentinel™ User’s Manual
4.2.8
Err 7: DSP Signal Error
4.2.9
Err 8: Temp Input Error
Problem:
This message indicates a temperature input error.
Cause:
The temperature exceeds the specified limits for the analog/RTD inputs option card or no input device is
connected.
Action:
Check the temperature transmitter and the connecting cable. Refer to
Diagnostics
in Chapter 2,
Initial
Setup
, and recalibrate the analog/RTD inputs option card.
4.2.10 Err 9: Press In Error
Problem:
This message indicates a pressure input error.
Cause:
The pressure exceeds the specified limits for the analog input card, or no input device is connected.
Action:
Check the pressure transmitter and the connecting cable. Refer to
“Entering Temperature and Pressure
and calibrate the analog input card.
4.2.11 Err 10: Totalizer Overflow
Problem:
The totalizers are unable to keep up with the total accumulated flow signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......