Chapter 7. Specifications
118
Sentinel™ User’s Manual
7.2 Electronics Specifications (cont.)
Modbus Map
The IGM878 Flow Transmitter, when equipped with the optional Modbus output card, can transmit flow data and
diagnostic information to a flow computer or SCADA, serially, using a Gould-type RTU protocol. For security and
audit-trail purposes, the unit must be programmed through the PanaView interface. This means that only the
Modbus function command
3
(read multiple registers), is valid.
Communication parameters: 9600, n, 8, 1
The format for the data exchange is as follows:
•
The
send
command (initiated by host flow computer or controller) is of the form:
<time delimiter> <Addr> <3> <First Register MSB>
<First Register LSB> <Register Count MSB>
<Register Count LSB> <CRC Low> <CRC High> <time delimiter>
•
The
response
(initiated by host flow computer or controller) is of the form:
[time delimiter] <Addr> <3> <Byte count> < Data ……… >
<CRC Low> <CRC High> <time delimiter>
The format for the returned data types is as follows:
Table 13 on page 109
is the data map for the IGM878 and Sentinel. The refresh rate indicates how often the
central controller updates the memory map, available using the ModBus port. The most time-critical information
is stored at the top of the register. This limits how deep the user has to go into the resister stack to gather the
real-time information. All values are IEEE format MSB first (big-endian).
Integer
Integer (I)
1 Register
Floating Point
Single Precision (FP)
2 Registers
Double Precision
Double Precision (DP)
4 Registers
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......