Chapter 6. Transducer Replacement
102
Sentinel™ User’s Manual
Note:
To replace T11 transducers in a pressurized pipe, see
“Replacing T11 Transducers in a Pressurized Pipe” on
.
The following equipment is required to complete the removal and re-installation of T11 transducers in a depressurized
pipe line:
• Torque wrench
• Spray lubricant
• Anti-seize lubricant
6.3.1
Removing the Old Transducer from a Depressurized Pipe
To remove the old transducer, complete the following steps:
1.
Remove the cover from the T11 conduit box, disconnect the BNC connector from the transducer, and remove the
conduit box.
2.
Remove the four bolts from the transducer sensor flange with a 9/16” socket wrench.
IMPORTANT:
Inspect the bolts to verify that they are not damaged. If one or more bolts need to be replaced, use only
approved 3/8-16 UNC x 1.5” SAE class 8 bolts.
3.
Remove the sensor flange from the transducer.
4.
Remove the transducer from the holder.
WARNING!
Follow all applicable safety codes while performing the following procedure.
WARNING!
Make sure the power is disconnected before performing the following steps.
WARNING!
Make sure the process line is shut down before removing transducers according to this
procedure. Serious injury or death may occur if the transducers are removed from a live
process line.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......