Sentinel™ User’s Manual
93
Chapter 4. Error Codes
4.2.4
Err 3: Velocity Range
Problem:
The velocity exceeds the limits programmed in the New Meter Browser information setup (see Chapter 2,
Initial Setup
).
Cause:
This error may be caused by the entry of improper programming data or by poor flow conditions and/or
excessive turbulence.
Action:
Make sure the actual flow rate is within the programmed limits. See Chapter 2,
Initial Setup
for details.
Refer to Chapter 5,
Diagnostics
, to correct any flowcell and/or transducer problems.
4.2.5
Err 4: Signal Quality
Problem:
The signal quality is outside the limits programmed in the New Meter Browser information setup (see
Chapter 2,
Initial Setup
).
Cause:
The peak of the upstream or downstream correlation signals has fallen below the correlation peak limit,
as set in the New Meter Browser information setup. This may be caused by a flowcell or electrical
problem.
Action:
Check for sources of electrical interference and verify the integrity of the electronics console by
temporarily substituting a test flowcell that is known to be good. Check the transducers and relocate
them, if necessary. See Chapter 5,
Diagnostics
, for instructions.
4.2.6
Err 5: Amplitude Error
Problem:
The signal amplitude exceeds the limits programmed in the New Meter Browser information setup (see
Chapter 2,
Initial Setup
).
Cause:
Solid or liquid particulates may be present in the flowcell. Poor coupling for clamp-on transducers.
Action:
Refer to Chapter 5,
Diagnostics
, to correct any flowcell problems.
4.2.7
Err 6: Cycle Skip, Accel.
Problem:
The acceleration exceeds the limits programmed in the New Meter Browser information setup (see
Chapter 2,
Initial Setup
).
Cause:
This condition is usually caused by poor flow conditions or improper transducer alignment.
Action:
Refer to Chapter 5,
Diagnostics
, to correct any flowcell and/or transducer problems.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......