Chapter 6. Transducer Replacement
110
Sentinel™ User’s Manual
6.4.1 Removing the Old Transducer from a Pressurized Pipe (cont.)
7.
Using the hydraulic pump valve, slowly release pressure from the hydraulic ram until the transducer has
retracted to the positive stop in the mechanism (see
below).
Figure 11: Transducer Retracted
8.
Close the ball valve and secure it in the closed position. Lock the ball valve in place using an appropriate
LOTO
(lock-out, tag-out) device and tags.
9.
Open the pressure relief valve to release the pressure of any gas trapped in the holder, then close the pressure
relief valve and torque it to 15 ft-lb.
10.
Disconnect the hydraulic pump and remove the insertion mechanism.
11.
Remove the T11 transducer assembly from the pipe and then remove the pusher and the 4-bolt sensor flange
from the transducer.
The T11 transducer removal procedure is now complete.
Transducer
Retracted
Danger Zone!
Keep hands clear as
much as is practical.
Ball Valve
(closed)
Relief Valve
Positive Stop
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......