Chapter 3. Operation
64
Sentinel™ User’s Manual
3.2.2 Function Verification Procedures
Before continuing with function verification, the flowmeter body must be installed in the pipeline and pressurized with
process gas. Also, the Sentinel's communications and outputs must be wired to the flow computer or plant
computer. Temperature and pressure sensors should be connected to either the Sentinel or the flow computer. The
Sentinel
must be powered for at least one minute to allow for self-test and initialization, as follows:
1.
Verify that the on-site control computer is not indicating fault condition from the Sentinel if the alarm output is
connected.
2.
If the alarm output is not connected, verify that the Sentinel is not reporting any errors by observing its display
and fault indicator (red) light beneath the display.
3.
Verify that the flow reading reported by the on-site control computer is reasonable and stable.
If none of the above procedures indicate a problem, the function verification is complete. If a problem is indicated,
proceed to
“Installation Troubleshooting Procedures”
below.
3.2.3 Installation Troubleshooting Procedures
The purpose of these procedures is to troubleshoot the Sentinel installation, if a fault condition was indicated while
performing the “
Function Verification Procedures”
in the previous section.
If the Sentinel failed to power on, check the wiring for the presence of power in a safe manner. If the power is wired
correctly and power is present, call Panametrics for assistance.
If the Sentinel is indicating a fault condition on the display or alarm output, refer to
Error Codes
, Chapter 4, or
Diagnostics
, Chapter 5 in this manual.
If the on-site control computer is not receiving flow data and the Sentinel display is not indicating a fault, check the
wiring of the outputs from the Sentinel to the on-site control computer or the flow computer, if used.
Check the programming of the flow computer, if used, and verify that its input and output wiring is correct. If you
cannot resolve the problem, call Panametrics for assistance.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......