TransPort™ PT878GC User’s Manual
157
Chapter 9.
3.
The next prompt, the transmitter sample size, is the number of pulses each transmitter (upstream and
downstream) emits. It is set to 8 by default.
Note:
Consult Panametrics Sensing before performing this step.
a.
Press
[ENTER]
to open the drop-down list.
b.
Use the [
] or [
] arrow keys to scroll to the recommended number.
c.
Press
[ENTER]
to confirm your selection.
4.
The next prompt asks for the zero cutoff. Near “zero” flow, the PT878GC may have fluctuating readings due to
small offsets (caused by factors such as thermal drift in the fluid). The zero cutoff causes velocity measurements
less than the cutoff to be reported as zero. To set the cutoff:
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your entry.
5.
The next prompt asks for velocity averaging, the number of velocity measurements to average together to
smooth out noise in the system.
a.
Press
[ENTER]
to open the drop-down list.
b.
Use the [
] or [
] arrow keys to scroll to the desired number.
c.
Press
[ENTER]
to confirm your selection.
6.
The final prompt, errors allowed, specifies the number of errors the meter can record before displaying an error
message.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your entry.
You have finished entering data in the Signal Parameter window. To leave this window,
•
Press the [
] key to return to the Signal Parameter tab, and the [
] arrow key to move to the Measurement
Mode tab.
•
Press
[F2]
(Cancel) to return to Operate Mode without confirming the changes.
•
Press
[F3]
(OK) to confirm the new data and return to Operate Mode.
Summary of Contents for PT878GC
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii TransPort PT878GC User s Manual...
Page 16: ...Chapter 2 6 TransPort PT878GC User s Manual Figure 3 Connection Locations...
Page 50: ...Chapter 3 40 TransPort PT878GC User s Manual...
Page 98: ...Chapter 5 88 TransPort PT878GC User s Manual...
Page 148: ...Chapter 8 138 TransPort PT878GC User s Manual...
Page 152: ...Chapter 9 142 TransPort PT878GC User s Manual Figure 131 Printout of a Typical Drive Report...
Page 190: ...Chapter 10 180 TransPort PT878GC User s Manual...
Page 196: ...Chapter 11 186 TransPort PT878GC User s Manual...
Page 198: ...Appendix A 188 TransPort PT878GC User s Manual...
Page 199: ...Appendix A TransPort PT878GC User s Manual 189 Figure 163 The Site Menu...
Page 201: ...Appendix A TransPort PT878GC User s Manual 191 Figure 165 The Program Menu Pipe Option...
Page 204: ...Appendix A TransPort PT878GC User s Manual 194 Figure 168 The Meter Menu...
Page 205: ...Appendix A TransPort PT878GC User s Manual 195 Figure 1 Logging Mneu...
Page 212: ...Appendix D 202 TransPort PT878GC User s Manual...
Page 218: ...Index 208 TransPort PT878GC User s Manual...
Page 220: ...2 Doc Tittle Manual Type...
Page 222: ...Warranty 4 TransPort PT878GC User s Manual no content intended for this page...
Page 223: ......