DigitalFlow™ XGF868i Service Manual
49
Appendix B. Calibration and Testing with PanaView
B.5.3.1 EPROM Test
To test the EPROM from the
Calibration/Test
menu:
1.
From the left pane of the
Calibration/Test
window, double-click on
Hardware Test
.
2. In the center pane, double-click on
EPROM TEST
.
3. The right pane displays the
EPROM Sum Value
. Click
[Next Item]
or
[Exit Page]
to return to the
Hardware Test
option,
and
[Exit Page]
again to return to the main
Calibration/Test
menu.
B.5.3.2 NVR Test
1.
From the left pane of the
Calibration/Test
window, double-click on
Hardware Test
.
Note:
If you are in the Hardware Test option, skip to step 2.
2. In the center pane, double-click on
NVR TEST
.
3. A message in the right pane cautions that the test will take several minutes, and that interrupting the test may
corrupt the memory. Click
[Next Item]
to continue the test.
4. If the NVR passes the test, the screen appears similar to
[Exit Page]
to return to the
Hardware
Test
option, and
[Exit Page]
again to return to the main
Calibration/Test
menu.